General Specifications |
|
Maximum Voltage Applied to any
Terminal |
600 Volts AC RMS or DC
|
|
Storage Temperature
|
-40 °C
to 60 °C (-40
°F to 140 °F) |
|
Operating Temperature |
-20 °C
to 55 °C (-4
°F to 131 °F) |
|
Temperature Coefficient
|
0.05 x (specified accuracy) per
°C for temperatures
< 18 °C or > 28
°C (< 64 °F
or > 82 °F) |
|
Relative Humidity |
|
0% to 95% @ 10
°C to 30
°C (50
°F to 86 °F) |
|
0% to 75% @ 30
°C to 40
°C (86
°F to 104 °F) |
|
0% to 40% @ 40
°C to 55
°C (104
°F to 131 °F) |
|
|
Vibration |
Random, 2g, 5-500 Hz per MIL-PRF-
28800F, Class 2 Instrument |
|
Shock |
1 meter drop per IEC 61010-1 2nd
Edition (1 meter drop test, six sides, oak floor) |
|
Electromagnetic Compatibility |
In an RF field of 3 V/M, accuracy
= specified accuracy
(EN 61326-1:1997) |
Safety
|
Complies with ANSI/ISA 82.02.01
(61010-1) 2004, CAN/CSA-C22.2 No. 61010-1-04, and IEC/EN
61010-1 2nd Edition for measurement category IV 600V (CAT IV) |
| Certifications |
CSA per standard CSA/CAN C22.2 No.
61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition |
| Batteries |
Four 'AA' batteries (NEDA 15A or
IEC LR6) |
| Battery Life |
| |
Insulation Test:
|
| |
|
| |
Resistance Measurements:
|
|
|
|
|
Tester can perform at least
1000 insulation tests with fresh alkaline batteries at
room temperature. These are standard tests of 1000V into
1MΩ with a duty cycle of 5 seconds
on and 25 seconds off. |
|
|
|
Tester can perform at least
2500 earth bond resistance measurements with fresh
alkaline batteries at room temperature. These are
standard tests of 1Ω with a duty
cycle of 5 seconds on and 25 seconds off. |
|
Size
|
5.00cm (H) x 10.0cm (W) x 20.3cm
(L)
1.97" (H) x 3.94" (W) x 8.00" (L) |
| Weight |
550gms (1.2lb) |
| IP Rating |
IP40 |
| Altitude |
| |
Operating: |
| |
Non-operating (storage): |
|
|
|
|
2000 meters CAT IV 600V,
3000 meters CAT III 600V |
|
12,000 meters |
|
| Over-range Capability |
110% of range |