Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

 
 Fluke Test Instruments Logo

Models 1523 and 1524 Reference Thermometer - Sensor Accuracies

 
Accuracies with Selected Probes (1523/1524)

Value at

Accuracy

Probe Type Number

-200°C

±0.014°C

5616-12

±0.025°C

5615-6

±0.027°C

5627A-12

n/a

5610-9

0°C

±0.021°C

5616-12

±0.021°C

5615-6

±0.049°C

5627A-12

±0.009°C

5610-9

100°C

±0.027°C

5616-12

±0.028°C

5615-6

±0.065°C

5627A-12

±0.009°C

5610-9

300°C

±0.040°C

5616-12

±0.043°C

5615-6

±0.103°C

5627A-12

n/a

5610-9

420°C

±0.050°C

5616-12

n/a

5615-6

±0.130°C

5627A-12

n/a

5610-9

 
Note: Includes readout accuracy, probe calibration, and probe drift.
 
PRT equivalent Temperature Accuracy (Readout Only)

Value at

Accuracy

-100°C

±0.011°C

0°C

±0.015°C

100°C

±0.019°C

200°C

±0.023°C

400°C

±0.031°C

600°C

±0.039°C

 
 
Thermistor equivalent Temperature Accuracy (Readout Only)

Value at

Accuracy

0°C

±0.002°C

25°C

±0.003°C

50°C

±0.006°C

75°C

±0.014°C

100°C

±0.030°C

 
 
Thermocouple Equivalent Temperature Accuracies (Readout Only)

Type 'B'

±0.85°C

from 600°C to 800°C

±0.68°C

from 800°C to 1000°C

±0.57°C

from 1000°C to 1800°C

Type 'C'

±0.32°C

from 100°C to 550°C

±0.71°C

from 550°C to 2300°C

Type 'E'

±0.52°C

from -200°C to 0°C

±0.22°C

from 0°C to 950°C

Type 'J'

±0.52°C

from -200°C to 0°C

±0.23°C

from 0°C to 1200°C

Type 'K'

±0.61°C

from -200°C to 0°C

±0.24°C

from 0°C to 1370°C

Type 'L'

±0.36°C

from -200°C to 0°C

±0.23°C

from 0°C to 1370°C

Type 'M'

±0.26°C

from -20°C to 0°C

±0.25°C

from 0°C to 400°C

±0.22°C

from 400°C to 1400°C

Type 'N'

±0.72°C

from -200°C to 0°C

±0.28°C

from 0°C to 1300°C

Type 'R'

±1.09°C

from -20°C to 0°C

±0.97°C

from 0°C to 500°C

±0.49°C

from 500°C to 1750°C

Type 'S'

±1.05°C

from -20°C to 0°C

±0.95°C

from 0°C to 500°C

 

±0.56°C

from 500°C to 1750°C

Type 'T'

±0.60°C

from -200°C to 0°C

±0.25°C

from 0°C to 400°C

Type 'U'

±0.54°C

from -200°C to 0°C

±0.24°C

from 0°C to 400°C

 
Note: Accuracies are based on internal Reference Junction Compensation. Refer to Technical manual for equivalent accuracies with an external reference junction.

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |