| Voltage DC |
|
Maximum Voltage: |
1000V |
|
Accuracy 1577: |
±(0.2%+2) |
|
Accuracy 1587: |
±(0.09%+2) |
| Accuracy 1587T: |
±(0.09%+2) |
|
Maximum Resolution: |
0.001V |
|
| Voltage AC (True-rms) |
| Maximum Voltage: |
1000V |
| Accuracy 1577: |
±(2%+3) |
| Accuracy 1587: |
±(2%+3) |
| Accuracy 1587T: |
±(2%+3) |
| AC Bandwidth 1587: |
5kHz* |
| Note* with low pass
filter: |
3db @ 800Hz |
| Maximum Resolution: |
0.1mV |
|
| Current DC |
| Maximum Current: |
400mA |
| Accuracy 1577: |
±(1.0%+2) |
| Accuracy 1587: |
±(0.2%+2) |
| Accuracy 1587T: |
±(0.2%+2) |
| Maximum Resolution: |
0.01mA |
|
| Current AC (True-rms) |
| Maximum Current: |
400mA |
| Accuracy 1577: |
±(2.0%+2)* |
| Accuracy 1587: |
±(1.5%+2)* |
| Accuracy 1587T: |
±(1.5%+2)* |
| Note* 1kHz bandwidth |
|
| Maximum Resolution: |
0.01mA |
|
| Resistance |
| Maximum Resistance: |
50MΩ |
| Accuracy 1577: |
±(1.2%+2) |
| Accuracy 1587: |
±(0.9%+2) |
| Accuracy 1587T: |
±(0.9%+2) |
| Maximum Resolution: |
0.1Ω |
|
| Capacitance |
1587/1587T only |
| Maximum Capacitance: |
9.9999µF |
| Accuracy: |
±(1.2%+2) |
| Maximum Resolution: |
1nF |
|
| Frequency |
1587/1587T only |
| Maximum Frequency: |
100kHz |
| Accuracy: |
±(0.1%+1) |
| Maximum Resolution: |
0.01Hz |
|
| Temperature Measurement |
1587/1587T only |
| Temperature: |
-40°C to 537°C |
| (excluding probe) |
-40°F to 998°F |
|
| Diode Test |
1587/1587T only |
| Range: |
6V |
| Accuracy: |
±(2%+1) |
| Resolution: |
1mV |
|
|
|
|
|
|
|
Insulation
Test |
|
Measurement |
1577 |
|
Range |
1587 |
|
|
1587T |
|
|
0.1MΩ
to 600MΩ |
|
0.1MΩ
to 2GΩ |
|
0.1MΩ
to 100MΩ |
|
|
Test
Voltages |
1577 |
|
|
1587 |
|
|
1587T |
|
|
500V, 1000V |
|
50V, 100V,
250V, 500V, 1000V |
|
50V, 100V |
|
|
Test
Voltage |
1577 |
|
Accuracy |
1587 |
|
|
1587T |
|
|
+20% -0% |
|
+20% -0% |
|
+20% -0% |
|
|
Insulation
Test |
1577 |
|
Current |
1587 |
|
|
1587T |
|
|
1mA nominal |
|
1mA nominal |
|
1mA nominal |
|
|
Auto
Discharge |
1577 |
|
|
1587 |
|
|
1587T |
|
|
Discharge
time <0.5sec for C=1µF or less |
|
Discharge
time <0.5sec for C=1µ or less |
|
Discharge
time <0.5sec for C=1µF or less |
|
|
Maximum |
1577 |
|
Capacitance Load |
1587 |
|
|
1587T |
|
|
Up to 1µF
load |
|
Up to 1µF
load |
|
Up to 1µF
load |
|
|
|
|
|
|
|
Environmental Specifications |
|
Operating
Temperature |
-20°C
to +55°C |
|
Storage
Temperature |
-40°C
to +60°C |
|
Humidity
(without Condensation) |
|
0% to 95% |
10°C
to +30°C |
|
0% to 75% |
30°C
to +40°C |
|
0% to 40% |
40°C
to +55°C |
|
|
Operating
Altitude |
2000 meters |
|
|
|
|
|
|
Safety
Standards |
|
Overvoltage
Category
|
Complies with
ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04,
and IEC/EN 61010-1 2nd Edition for measurement CAT III 1000V and
CAT IV 600V. |
|
Agency Approvals
|
2000 m CAT III
1000V, CAT IV 600V
3000 m CAT II 1000V, CAT III 600V |
|
|
|
|
|
|
General
Specifications |
|
Size |
203mm x 100mm x
50mm (with holster) |
|
Weight |
624g |
|
Warranty |
3 years |
|
|
|
Meter use: |
1000 hours |
|
Insulation |
Meter can
perform at least 1000 insulation tests |
|
Test use: |
with fresh
alkaline batteries at room temperature. |
|
|
These are
standard tests of 1000V into 1MΩ with |
|
|
a duty cycle
of 5 seconds on and 25 seconds off. |
|
|
IP Rating |
IP40 |
|
|
|