Bus Health Technical
Specifications |
|
Bus Health Test |
Verifies the electrical
parameters of industrial bus systems using automatic
measurement and analysis functions. Next to that, an
Eyepattern mode is provided for visual inspection of signal
quality. |
|
Parameter Classification
Default Values
|
Well
within limits + 'good', gives green tick mark.
Within specified percentage of the limit values = 'weak',
gives orange
tick-mark.
Beyond limit values = 'bad', gives red tick mark.
Note: Limits values, as a default, are based on industry
standard for the selected bus type or may be set by the
user. The tolerance band (percentage) taken into account
for the classification 'weak' may be altered by the user. |
|
Bus System Supported
|
AS-i (EN
50295, 168 kb/s);
CAN-bus (ISO-11898, up to 1 Mb/s);
Modbus 232 (EIA-232 up to 115 kb/s);
Modbus 485 (EIA-485 up to 10Mb/s);
Foundation Fieldbus H1 (61158 type 1, 31.25 kb/s);
Profibus DP (EIA-485 up to 12Mb/s);
Profibus PA (61158 type 1, 31.25 kb/s);
Ethernet 10Base2 (coaxial, 10 Mb/s);
Ethernet 10BaseT (UTP, 10 Mb/s);
Ethernet 100Base2 (100 Mb/s);
RS-232 (EIA-232, up to 115 kb/s);
RS-485 (EIA-485, up to 10 Mb/s). |
|
Modes
|
Waveform
Parameter analysis with automatic read-out and parameter
validation (good/weak/bad); test limits are based on
industry
standards or may be set by the user.
Eyepattern display mode. |
|
Measured Parameters (where
applicable) |
bias
voltage level
Signal amplitude
pulse width or baud rate
rise-time
fall time
jitter
signal distortion
noise HF
Noise LF
in-band noise |
|
Eyepattern Display Mode |
Gives waveform display of bus
signals with user selectable persistence of the display.
Time-base, attenuator and persistence settings may be
altered by the user. |
|
Persistence Modes |
off - short - medium - long -
infinite |
|
| |
| |
|
Oscilloscope
Specifications |
|
Description |
|
Bandwidth |
|
Max. Real Time Sample
Rate |
|
|
Model 215C |
Model 225C |
|
100 MHz |
200 MHz |
|
1 GS/s |
2.5 GS/s |
|
|
Input Sensitivity |
2 mV to 100 V/div |
|
Time-base range |
5 ns/div to 2 min/div |
|
Inputs and digitizers |
2 inputs per model plus
trigger/DMM input |
|
Independently Floating
Isolated Inputs |
Up to 1000 volts between inputs,
references and ground |
|
Trigger Types
|
Connect-and-View, Free Run,
Single Shot, Edge, Delay, Video, Selected pulse width and
External Dual slope trigger and Event trigger (n-cycle) |
|
Glitch Capture
|
up to 3 ns using Pulse Width
triggering; 50 ns peak detected at 5 us/div to 1 min/div |
|
Memory
|
|
|
Display |
144mm full color LCD with fast
update rate |
|
Persistence |
Digital persistence giving
analog oscilloscope-like waveform decay |
|
Waveform Compare |
Waveform reference with
automatic Pass/Fail testing |
|
Max. Record
Length/Scope Mode |
|
Max in Scope Record
Mode |
|
|
3,000 points per input |
|
27,500 points per input or
more (5ms/div ...2min/div) |
|
|
Scope Measurements |
7 cursor + 30 automatic,
including power and Vpwm |
|
True RMS Multimeter |
5000 counts, single input |
|
| |
| |
|
Environmental
Specifications |
|
Operating Temperature |
0ฐC
to 50ฐC |
|
Storage Temperature |
-20ฐC
to +60ฐC |
|
Operating Altitude |
3000 meters |
|
| |
| |
|
Safety |
|
Electrical Safety |
1,000V CAT II, 600V CAT III (EN
61010-1) |
|
| |
| |
|
General Specifications |
|
Line Power |
Adapter/Battery/charger included |
|
Battery Power |
4 hours NiMH |
|
Dimensions |
256mm (L) x 169mm (W) x 64mm (D) |
|
Weight |
2 kg |
|
Warranty |
3-years on main instrument,
1-year on accessories |
|
|