Models 4180 and 4181 Precision
Infrared Calibrator - Specifications |
|
|
|
|
Item |
Model 4180 |
Model 4181 |
| Temperature range @ 23°C
ambient 0.95 emissivity |
-15°C
to 120°C |
35°C
to 500°C |
| Display Accuracy
Note 1. |
±0.40°C at
-15°C |
±0.35°C at
35°C |
| ±0.40°C at
0°C |
±0.50°C at
100°C |
| ±0.50°C at
50°C |
±0.70°C at
200°C |
| ±0.50°C at
100°C |
±1.20°C at
350°C |
| ±0.55°C at
120°C |
±1.60°C at
500°C |
| Stability |
±0.10°C at
-15°C |
±0.50°C at
35°C |
| ±0.05°C at
0°C |
±0.20°C at
200°C |
| ±0.10°C at
120°C |
±0.40°C at
500°C |
| Uniformity (5.0
inch Dia of center of target) Note 2. |
±0.15°C at
-15°C |
±0.10°C at
35°C |
| ±0.10°C at
0°C |
±0.50°C at
250°C |
| ±0.25°C at
120°C |
±1.00°C at
500°C |
| Uniformity (2.0
inch Dia of center of target) Note 2. |
±0.10°C at
-15°C |
±0.10°C at
35°C |
| ±0.10°C at
0°C |
±0.25°C at
250°C |
| ±0.20°C at
120°C |
±0.50°C at
500°C |
| Heating Time |
15 min: -15°C
to 120°C |
20 min: 35°C
to 500°C |
| 14 min: 23°C
to 120°C |
|
| Cooling Time |
15 min: 120°C
to 23°C |
40 min: 500°C
to 100°C |
| 20 min: 23°C
to -15°C |
|
| Stabilization Time |
10 minutes |
10 minutes |
| Nominal Emissivity Note 3. |
0.95 |
0.95 |
| Thermometer Emissivity
Compensation |
0.9 to 1.0 |
0.9 to 1.0 |
| Target Diameter |
152.4mm (6 inches) |
152.4mm (6 inches) |
| Computer Interface |
RS-232 |
RS-232 |
| Power (50/60 Hz,
630 W) |
115V AC (±10%),
6.3A |
|
| 230V AC (±10%),
3.15A |
| Power (50/60 Hz,
1000 W) |
|
115V AC (±10%),
10A |
| 230V AC (±10%),
5A |
|
|
6.3A 250V slow blow |
10A 250V slow blow |
| 3.15A 250V |
5A 250V very fast blow |
| Dimensions |
356mm (H) x 241mm (W) x 216mm (D) |
| Weight |
9.1kg |
9.5kg |
| Safety |
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
|
|
|
| Note 1. |
For 8um to 14um spectral band thermometers with emissivity set
between 0.9 and 1.0 |
| Note 2. |
The uniformity specification refers to how IR thermometers with
different spot sizes both focused at the |
| |
center of the target will measure the same temperature. |
| Note 3. |
The target has a nominal emissivity of 0.95, however it is
radiometrically calibrated to minimize emissivity |
| |
related uncertainties. |
|
|
|