Technical Specifications |
|
Input
Characteristics |
Input
impedance: 1 MW, 20 pF
Voltage Rating: 600V rms, CAT III |
|
Volt/Amps/Hertz Display |
True-rms voltage (ac+dc)
Ranges: 5.000V, 50.00V, 500.0V, 1250V*
Accuracy: ±(1% +10 counts)
True-rms current (ac + dc)
Ranges: 50.00A, 500.0A, 5.000 kA, 50.00 kA
Accuracy: ±(1% +10 counts)
Frequency
Ranges: 10.0 Hz to 15.0 KHz
Accuracy: ±(0.5% + 2 counts)
Crest Factor
Ranges: 1.0 to 10.0
Accuracy: ±(5% + 1 count) |
|
Power |
Watts,
VA, VAR Reactive Power 1-phase and 3-phase, 3 conductor
balanced loads
Ranges: 250 W, 2.50 kW, 25.0 kW, 250 kW, 2.50 MW,
25 MW, 250 MW, 625 MW, 1.56 GW
Accuracy: ± (2% + 6 counts) Total Power,
± (4% + 4 counts) Fundamental power.
PF Power Factor
Ranges: 0.00 to 1.00
Accuracy: ± 0.04
DPF Displacement Power Factor
Range: 0.00 to 0.25 Not specified,
0.25 to 0.90 ± 0.04, 0.90 to 1.00 ± 0.03
Hz frequency fundamental
Range: 40.0 to 70.0 Hz
Accuracy: ± (0.5% + 2 counts) |
|
Harmonics |
Volts,
Amps, Watts:
Fundamental:
V,A ± (3% + 2 counts), W ± (5% + 2 counts)
Range: 2 to 31st Harmonic
Accuracy: V,A ± (5% + 3 counts), W ± (10% + 10 counts)
Range: 32 to 51st Harmonic
Accuracy: V,A ± (15% + 5 counts), W ± (30% + 5 counts)
Frequency of fundamental:
Range: 40 Hz to 70 Hz
Accuracy: ± 0.25 Hz
Phase: Volt & Amps (between Fund & Harmonic) Accuracy:
2nd (± 3º) . . . 51st (± 15º)
Watts (between Volt Fund. & Amps Harmonic)
Accuracy: Fund (± 15º)
K-Factor (Amps & Watts):
Range: 1.0 to 30.0
Accuracy: ± 10%
THD:
Range: 0.00 to 99.99
Accuracy ± (3% + 8 counts) |
|
Sags
and Swells Recording |
Recording times (selectable): 4 min to 16 days
Vrms actual, Vrms max, min (AC+DC):
5.000 V, 50.00 V, 500.0 V, 1250 V*
Accuracy: Readings ±(2% + 10 counts), Cursor readings ±
(2% + 12 counts), Cursor Readings Average ± (2% + 10
counts)
Arms actual, Arms max, min (AC+DC):
50.00 A, 500.0 A, 5.000 kA, 50.00 kA |
|
Recording |
Recording times (selectable) : 4 min to 16 days
Parameters: Choose up to two from one of the groups
below:
V/A/Hz: Line voltage, current, frequency
Power: Watts, VA, VAR, PF, DPF, frequency
Harmonics: THD, volts(Fund. & Harmonics), Amps (F&H)
Watts (F&H) Freq. (H),%(H) of total, Phase(H), KF
Ohms: Ohms, Diode, Continuity, Capacitance
Temperature: °C or °F
Scope: DC voltage, DC Current, AC Voltage, AC Current,
Frequency, Pulse Width + or -, Phase, Duty cycle + or -,
Peak max, Peak min, Peak min-max, Crest Factor. |
|
Transients |
Minimum
pulse width: 40 ns
Useful bandwidth input 1: DC to 1 MHz (with test leads
TL24)
Number of transients 40
Voltage threshold settings: 20%, 50%, 100%, 200% above
or below reference signal.
Reference signal: After START, the Vrms and frequency
of the signal are measured. From these data a pure
sinewave is calculated as reference for threshold setting.
Vpeak min, Vpeak max at cursor:
10 V, 25 V, 50 V, 125 V, 250 V, 500 V, 1250 V
Accuracy: ±5% of full scale. |
|
Inrush
Current |
Current
ranges (selectable):
1 A, 5 A, 10 A, 50 A, 100 A, 500 A, 1000 A
Inrush times (selectable):
1 s, 5 s, 10 s, 50 s, 100 s, 5 min
Cursor readings
A Peak max at cursor 1 and cursor 2
Accuracy: ± 5% of full scale
Time between cursors
4 to 235 pixels (1 pixel = inrush time/250)
Accuracy: ±(0.2% + 2 pixels) |
|
Scope,
dual channel scope with measurement reading |
Input
impedance
Input 1:
Ranges: 1 MW//12 pF; with BB
120: 20pF
Accuracy: ±2 pF; with BB120 ±3 pF
Input 2:
Ranges: 1 MW//10 pF; with BB
120: 18pF
Accuracy: ±2 pF; with BB120 ±3 pF
Vertical
Voltage Ranges:
Ranges: 50 mV/div to 500 V/div
Accuracy: ± (1% + 2 pixels)
Vertical sensitivity, resolution:
Ranges: 5 mV/div to 500 V/div, 8 bit (256 levels)
Bandwidth input 1 (voltage):
Ranges: DC to 20 MHz at inputs, or with BB120 and VPS
100-R probe (opt); 1 MHz with TL24 Leads
Bandwidth input 2 (current):
DC to 15 kHz at inputs
10 kHz with 80i-500s Current Clamp
Coupling:
DC, AC (10 Hz -3 dB)
Horizontal
TimeBase modes:
Normal, roll, single
TimeBase ranges:
Ranges: 60 s/div to 20 ns/div
Accuracy: ± (0.4% + 1 pixel)
Sampling rate: 25 MS/s
Record length (min/max samples):
512 per channel
Trigger source:
Ranges: Input 1 or Input 2 or Automatic selection
Trigger Mode:
Automatic Connect-and ViewTM,
Free Run, Single Shot.
Connect-and-ViewTM:
Advanced automatic triggering that recognizes signal
patterns and automaticaly adjusts triggering, timebase and
amplitude. Automatically displays stable pictures of
complex and dynamic signals like motor drive and control
signals.
Pre Trigger:
Up to 10 divisions
Measurement readings, per channel selectable:
Ranges: Volts & Amps (DC, AC, AC + DCrms, Peak max, Peak
min, Peak min/max). Frequency, Duty cycle + or =, Phase,
Pulse Width + or -, Crest factor. |
|
Ohms,
Diode, Continuity, Capacitance |
Ohms:
Ranges: 500.0 W, 5.000 kW,
50.00 kW, 500.0 kW,
5.000 MW, 30.00 MW
Accuracy: ±(0.6% + 5 counts)
Diode voltage:
Ranges: 0 to 3.000 V
Accuracy: ±(2% + 5 counts)
Continuity, shorts > 1 ms:
Ranges: Beeper on at < (30 W ±
5 W)
Accuracy: Not Applicable
Capacitance:
Ranges: 50.00 nF, 500.0 nF, 5.000 µF, 50.00 µF,
500.0 µF
Accuracy: ±(2% + 10 counts)
Temperature:**
Ranges: -100.0 ºC to 400.0 ºC, (-200.0 ºF to 800.0 ºF)
Accuracy: ±(0.5% + 5 counts)
Max current, max open circuit volt:
Ranges: 0.5 mA, < 4 V (all functions above) |
|
Memory |
Number of
screens: 20 |
|
Optically Isolated RS-232 Interface |
To
printer:
Supports HP Laserjet®, Deskjet®,
Epson FX/LQ and Postscript printers with optional PAC91
Printer Adapter Cable
To PC:
FlukeView® Power Quality Analyzer Software with
PM9080 Interface Adapter included. |
|
FlukeView Power Quality Software |
Hardware requirements:
PC or 100% compatible with Windows® 95, 98, Me,
2000, NT 4.0. |
|
*Rated 600V CAT III
**Requires optional temperature accessory |
|
|
|
General Specifications |
|
Power |
Line
voltage adapter/battery charger included
Installed battery: Rechargeable NiCd pack (4 to 6
Vdc)
Operating time: 4 hours
Charging time: 4 hours (Fluke 43B off) 12 Hours
(Fluke 43B on)
Refresh Cycle: 8 to 14 hours (to keep NiCd battery
capacity optimal) |
|
Environmental |
Temperature: 0ºC to 50ºC (32ºF to 122ºF)
Environmental: MIL 28800E, Type 3, Class III, Style
B
Enclosure: IP51 (dust, drip water proof) |
|
Mechanical Data |
Size:
(H x W x D) 232 x 115 x 50 mm (9.1 x 4.5 x 2 inches)
Weight: 1.1 kg (2.5 lbs) |
|
Safety |
For
measurements on 600 Vrms Category III installations,
Pollution Degree 2, in accordance with EN61010-1 (1993)
(IEC1010-1), ANSI/ISA S82.01-1994, CAN/CSA-C22.2 No.
1010.1-92 UL3111-1
Surge protection: 6 kV on input 1 and 2
Floating measurements: 600 Vrms from any terminal
to ground. |
|
Warranty |
3 years
parts and labor on Fluke 43B, 1 year on accessories. |
|
|
|