General
Specifications |
|
Warm Up Time |
Twice the time since last warmed up, to
a maximum of 30 minutes |
|
Setting Time |
Less than 5 seconds for all functions and ranges
except as noted |
|
Standard Interfaces |
RS-232 and Ethernet |
|
Temperature Performance |
Operating: 0°C to 50°C
Calibration (tcal): 15°C to 35°C
Storage: -20°C to +70°C |
|
Temperature Coefficient |
Temperature coefficient for temperatures outside
tcal ±5°C is 10% of stated specification per °C
for temperatures in the range of 0°C to 35°C. Above 35°C, the
temperature coefficient is 20% of the stated specification per °C. |
|
Relative Humidity |
Operating: <80% to 30°C, <70%
to 40°C, <40% to 50°C
Storage: <95%, non-condensing |
|
Altitude |
Operating: 2000 meters (6500ft) maximum
Non-Operating: 12200 meters (40000ft) maximum |
|
Safety |
Meets EN 61010-1-2001, CAN/CSA-C22.2 No. 61010-1-04.
UL 61010-1: 2004 |
|
Analog Low Isolation |
20V |
|
EMC |
Meets EN 613261: 2006 |
|
Line Power |
Line Voltage (selectable): 100V, 120V,
220V, 240V
Line Frequency: 47Hz to 63Hz
Line Voltage Variation: +10% about line voltage setting |
|
Power Consumption |
600VA |
|
Dimensions |
Height: 19.3cm (7.6 ins)
Width: 43.2cm (17 ins)
Width with Handles: 44.3cm (17.5 ins)
Depth: 53.8cm (21.2 ins) |
|
Weight (without options) |
22kg (48lb) |
|
Absolute Uncertainty Definition |
The specification includes stability,
temperature coefficient, linearity, line and load regulation, and
traceability of the external standards used for calibration. It is
not necessary to add anything to determine the total specification
for the temperature range indicated. |
|
Specification Confidence Interval |
99% |