Model 50 Series II Thermometers - Specifications

 
Technical Specifications
Measurement accuracy Above -100 °C (-148 °F):
J, K, T, E, and N-type: ±[0.05% + 0.3°C (0.5°F)]*
R and S-type : ±[0.05% + 0.4°C (0.7°F)]*

Below -100 °C (-148 °F):
J, K, E, and N-types: ±[0.20% + 0.3°C (0.5°F)]*
T-type: ±[0.50% + 0.3°C (0.5°F)]
Display resolution 0.1°C / °F / K < 1000°
1°C / °F / K ³ 1000°
Measurement range J-type: -210°C to 1200°C (-346°F to 2192°F)

K-type: -200°C to 1372°C (-328°F to 2501°F)

T-type: -250°C to 400°C (-418°F to 752°F)

E-type: -150°C to 1000°C (-238°F to 1832°F)

N-type: -200°C to 1300°C (-328°F to 2372°F)*

R and S-type: 0°C to 1767°C (32°F to 3212°F)*
Temperature scale ITS-90
Applicable standards NIST-175
 
Note: Only Models 53 Series II and 54 Series II thermometers are capable of measuring N, R and S-type thermocouples.
 
Environmental Specifications
Operating temperature -10°C to 50°C (14°F to 122°F)
Storage temperature -40°C to 60°C (-40°F to 140°F)
Humidity 0% to 90%; 0°C to 35°C (32°F to 95°F)
0% to 70%; 0°C to 50°C (32°F to 122°F)
 
 
Safety Specifications
Overvoltage Category CSA C22.2 No. 1010.1, 1992; EN 61010 Amendments 1,2
Agency Approvals CE, CSA, TUV (pending)
 
 
Mechanical and General Specifications
Dimensions 173mm (L) x 86mm (W) x 38mm (D)
Weight 400g
Batteries 3 AA batteries; typical 1000-hour life

 