Current
Measurement |
Range: 0 to
24 mA
Resolution: 0.001 mA
Accuracy: 0.015 % Rdg + 2 LSD |
|
Current Sourcing |
Range: 0 to
20 mA or 4 to 20 mA
Accuracy: 0.015 % Rdg + 2 LSD
Drive Capability: 700 W at 20 mA |
|
Current
Simulation |
Range: 0 to
20 mA or 4 to 20 mA
Accuracy: 0.015 % Rdg + 2 LSD |
|
Loop Power |
While measuring mA,
24 V |
|
Voltage
measurement |
Range: 0 to
28 V DC
Resolution: 1 mV
Accuracy: 0.015 % Rdg + 2 LSD |
|
|
|
|
|
|
Environmental Specifications |
|
Operating
Temperature |
-10 ºC to 55 ºC |
|
Storage
Temperature |
-40 ºC to 60 ºC |
|
Humidity (Without
Condensation) |
0 to 95 % (0 ºC to
30 ºC);
0 to 75 % (30 ºC to 40 ºC)
0 to 45 % (40 ºC to 50 ºC) |
|
Temperature
coefficient |
-10 ºC to 18 ºC, 28
ºC to 55 ºC,
±.005 % of range per ºC |
|
|
|
|
|
|
Safety Specifications |
|
Safety |
CSA C22.2 No.
1010.1:1992 |
|
EMC |
EN50082-1:1992 and
EN55022:1994 Class B |
|
Maximum voltage |
30 V (protection
circuit for 230 V) |
|
ATEX
certification |
Ex II G Eex ia IIC
T4 Zones 1 and 2,
ZELM 02 ATEX 0120 X |
|
FM certification |
N.I. Class 1,
Division 2, Groups A-D |
|
CE certification |
EN61010-1 and
EN61326 |
|
|
|
|
|
|
Mechanical and General
Specifications |
|
Size |
152 x 70 x 41 mm |
|
Weight |
224 g (350 g
including holster) |
|
Batteries |
One 9 V battery
Alkaline |
|
Warranty |
Three years |
|
Shock & Vibration |
per MIL-T-28800 for
a Class 2 Instrument |
|
Display current
and % of span |
Yes |
|
Auto step, auto
ramp |
Yes |
|
Span Check |