Model 719 Pressure Calibrator -
Specifications |
|
|
|
|
Technical Specifications |
|
Fluke 719-30G |
|
Range: |
-12 to 36 psi, -850 mbar to 2.4
bar |
|
Resolution: |
0.001 psi, 0.1 mbar |
|
Accuracy: |
0.025% of range for 6 months |
|
|
0.035% of range for 1-year |
|
|
Fluke 719-100G |
|
Range: |
-12 to 120 psi, -850 mbar to 8 bar |
|
Resolution: |
0.01 psi, 1 mbar |
|
Accuracy: |
0.025% of range for 6 months |
|
|
0.035% of range for 1-year |
|
|
mA |
|
Range: |
0 to 24 mA |
|
Resolution: |
0.001 mA |
|
Accuracy: |
.015% of reading + 2 counts |
|
Notes: |
Maximum load, 1000Ω,
750Ω in HART mode |
|
|
Loop Supply |
|
Range: |
24V dc |
|
Resolution: |
N/A |
|
Accuracy: |
±10% |
|
|
|
|
|
|
|
Pump Specifications |
|
Maximum Pressure Generation |
120 PSI (8 BAR) ( 100 PSI model) |
|
Pump to |
|
30 PSI in less than 15 seconds,
with 1 meter of 1/8 inch test hose |
|
100 PSI in less than 45 seconds,
with 1 meter of 1/8 inch test hose |
|
|
Pump Cycle between Rebuilds |
5000 cycles |
|
Total Pump Life |
20,000 calibration cycles |
|
|
|
|
|
|
General Specifications |
|
Storage Temperature |
-30°C to 60°C |
|
Operating Temperature |
-10°C to 55°C |
|
Relative Humidity |
|
-10°C
to 30°C |
95% |
|
30°C
to 40°C |
75% |
|
40°C
to 50°C |
45% |
|
50°C
to 55°C |
35% |
|
|
Shock |
1 meter drop test |
|
Safety |
CSA C22.2 No. 1010.1:1992 |
|
EMC |
EN 50082-1:1992 and EN 55022:1994 Class
B |
|
Dimensions
|
210mm (L) x 87mm (W) x 60mm (D)
8.28 ins (L) x 3.41 ins (W) x 2.19 ins (D) |
|
Weight |
912 gms (2 lbs) |
|
Power Supply |
Two 9 volt Alkaline batteries |
|
Battery Life |
12 hours with 12mA into 500Ω,
200 pump cycles to 100 psi,
500 pump cycles to 30 psi |
|
Warranty |
3-Years (1-Year for pump) |
|
|
|