Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

 Fluke Test Instruments Logo

Model 741B Documenting Process Calibrator - Specifications

 
Source Accuracy
Range (full scale) Accuracy (% of reading + % of full scale)
110.000 mV 0.01% + 0.005%
1.10000V 0.01% + 0.005%
15.0000V 0.01% + 0.005%
Source 22.000 mA 0.01% + 0.015%
Simulate 22.000 mA 0.02% + 0.03%
11.000W 0.01% + 20 mW
110.00W 0.01% + 40 mW
1.1000 kW 0.02% + 0.5MW
11.000 kW 0.03% + 5W
0.00 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 2 Hz
22.000 to 50.000 kHz 5 Hz
 
 
Measurement Accuracy
Range (Full scale) Accuracy (% of reading + % of full scale)
110.000 mV DC 0.025% + 0.015%
1.10000V DC 0.025% + 0.005%
11.0000V DC 0.025% + 0.005%
110.000V DC 0.05% + 0.005%
300.00V DC 0.05% + 0.005%
V AC, 20 to 40 Hz 2% + 10 counts
V AC, 40 to 500 Hz 0.5% + 5
V AC, 500 to 1 kHz 2% + 10
V AC, 1 kHz to 5 kHz 10% + 20
30.000 mA DC 0.01% + 0.015%
110.00 mA DC 0.01% + 0.015%
11.000W 0.05% + 50 mW
110.00W 0.05% + 50 mW
1.1000 kW 0.05% + 0.5W
11.000 kW 0.1% + 10W
1.00 to 109.99 Hz 0.05 Hz
110.0 to 1099.9 Hz 0.5 Hz
1.100 to 10.999 kHz 5 Hz
11.00 to 50.00 kHz 50 Hz
 
 
RTD's and Thermocouples
Device Measure Accuracy Source Accuracy

Notes

10Ω Cu (427) 3°C 1°C Note: Resolution: 0.1°C, except 1°C for 10Ω Cu. Best case, mid-range accuracies. Sensor inaccuracies not included. For 2-wire and 3-wire RTD measurements, add 0.4°C. Thermocouple accuracies with external cold junction. For internal junction add 0.2°C.
100Ω Pt (3916) 0.3°C 0.1°C
100Ω Pt (3926) 0.3°C 0.1°C
100Ω Pt (385) 0.3°C 0.1°C
200Ω Pt (385) 0.3°C 0.1°C
500Ω Pt (385) 0.3°C  0.1°C
1000Ω Pt (385) 0.3°C 0.1°C
120Ω Ni (672) 0.3°C 0.1°C
E 0.3°C 0.2°C
N 0.5°C 0.3°C
J 0.3°C 0.2°C
L 0.3°C 0.2°C
K 0.3°C 0.3°C
T 0.3°C 0.3°C
U 0.3°C 0.3°C
B 0.9°C 0.8°C
R 1.0°C 0.9°C
S 0.9°C 0.9°C
C 0.6°C 0.6°C
 
 
Technical Information
Environmental Specifications * Operating Temperature: -10ºC to 50ºC (-20ºC typical except for frequency and ac voltage measurement)
Storage Temperature: -20ºC to 60ºC
Operating Altitude: 2800 m above mean sea level (9186 ft)
Enclosure Protection: Designed to meet IEC529 IP52 (normal operating vacuum for dust)
Ramp Function Source functions: Voltage, current, resistance, frequency, temperature
Rate: 4 steps/second
Trip Detect: Continuity or voltage (continuity detection not available when sourcing current)
Loop Power Function Voltage: Selectable, 24V or 28V
Accuracy: 5%
Maximum current: 22 mA, short circuit protected
Maximum input voltage: 30V DC
Note*(all calibrator specifications apply from +18°C to +28°C unless stated otherwise)
 
Mechanical  and General Specifications
Dimensions 130 x 236 x 61 mm
Weight 1.4 kg (3 lb 1 oz)
Internal Battery Pack NiCd, 7.2V, 1700 mAh;
Battery Life Typical usage > eight hours
Battery Replacement Via snap-shut door without opening calibrator; no tools required
Side Port Connections
  • Pressure module connector
  • Connection for optional battery eliminator
Safety Complies with CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Data Storage Capacity 1 day of calibration results
90 Day Specifications The standard specification intervals for the 741B is 1 and 2 years. Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. Floor specifications, expressed as "% of f.s." or "counts" or "ohms" remain constant.

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |
 Site Map | Search | About Us | Contact Us

 

Rapid-Tech Equipment
ABN 79 883 184 102
1300 Ferntree Gully Road. Scoresby. Victoria 3179
P.O. Box 4243 Knox City Centre. Victoria 3152

Phone: +61 3 9763 3299     Fax: +61 3 9753 2132