Model 743B Documenting Process Calibrator -
Specifications |
|
|
Source Accuracy |
|
Range
(full scale) |
Accuracy (% of reading + % of full scale) |
|
110.000 mV |
0.01% + 0.005% |
|
1.10000V |
0.01% + 0.005% |
|
15.0000V |
0.01% + 0.005% |
|
Source 22.000 mA |
0.01% + 0.015% |
|
Simulate 22.000 mA |
0.02% + 0.03% |
|
11.000W |
0.01% + 20 mW |
|
110.00W |
0.01% + 40 mW |
|
1.1000 kW |
0.02% + 0.5W |
|
11.000 kW |
0.03% + 5W |
|
0.00 to 10.99 Hz |
0.01 Hz |
|
11.00 to 109.99 Hz |
0.1 Hz |
|
110.0 to 1099.9 Hz |
0.1 Hz |
|
1.100 to 21.999 kHz |
2 Hz |
|
22.000 to 50.000
kHz |
5 Hz |
|
|
|
|
|
|
Measurement Accuracy |
|
Range
(Full scale) |
Accuracy (% of reading + % of full scale) |
|
110.000 mV DC |
0.025% + 0.015% |
|
1.10000V DC |
0.025% + 0.005% |
|
11.0000V DC |
0.025% + 0.005% |
|
110.000V DC |
0.05% + 0.005% |
|
300.00V DC |
0.05% + 0.005% |
|
V AC, 20 to 40 Hz |
2% + 10 counts |
|
V AC, 40 to 500 Hz |
0.5% + 5 |
|
V AC, 500 to 1 kHz |
2% + 10 |
|
V AC, 1 kHz to 5
kHz |
10% + 20 |
|
30.000 mA DC |
0.01% + 0.015% |
|
110.00 mA DC |
0.01% + 0.015% |
|
11.000W |
0.05% + 50 mW |
|
110.00W |
0.05% + 50 mW |
|
1.1000 kW |
0.05% + 0.5W |
|
11.000 kW |
0.1% + 10W |
|
1.00 to 109.99 Hz |
0.05 Hz |
|
110.0 to 1099.9 Hz |
0.5 Hz |
|
1.100 to 10.999 kHz |
5 Hz |
|
11.00 to 50.00 kHz |
50 Hz |
|
|
|
|
|
|
RTD's and Thermocouples |
|
Device |
Measure Accuracy |
Source Accuracy |
Notes |
|
10Ω
Cu (427) |
3°C |
1°C |
Note*Resolution: 0.1°C,
except 1°C for 10Ω Cu. Best case, mid-range accuracies. Sensor
inaccuracies not included. For 2-wire and 3-wire RTD measurements, add
0.4°C. Thermocouple accuracies with external cold junction. For
internal junction add 0.2°C. |
|
100Ω
Pt (3916) |
0.3°C |
0.1°C |
|
100Ω
Pt (3926) |
0.3°C |
0.1°C |
|
100Ω
Pt (385) |
0.3°C |
0.1°C |
|
200Ω
Pt (385) |
0.3°C |
0.1°C |
|
500Ω
Pt (385) |
0.3°C |
0.1°C |
|
1000Ω
Pt (385) |
0.3°C |
0.1°C |
|
120Ω
Ni (672) |
0.3°C |
0.1°C |
|
E |
0.3°C |
0.2°C |
|
N |
0.5°C |
0.3°C |
|
J |
0.3°C |
0.2°C |
|
L |
0.3°C |
0.2°C |
|
K |
0.3°C |
0.3°C |
|
T |
0.3°C |
0.3°C |
|
U |
0.3°C |
0.3°C |
|
B |
0.9°C |
0.8°C |
|
R |
1.0°C |
0.9°C |
|
S |
0.9°C |
0.9°C |
|
C |
0.6°C |
0.6°C |
|
|
|
|
|
|
Mechanical and General Specifications |
|
Dimensions |
130 x 236 x 61 mm |
|
Weight |
1.4 kg (3 lb 1 oz) |
|
Internal Battery
Pack |
NiCd, 7.2V, 1700 mAh |
|
Battery Life |
Typical usage > eight
hours |
|
Battery Replacement |
Via snap-shut door
without opening calibrator; no tools required |
|
Side Port
Connections |
- Pressure module connector
- RS-232 connector to interface to your PC
- Connection for optional battery eliminator
|
|
Safety |
Complies with CAN/CSA
C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
|
Data Storage
Capacity |
1 week of calibration
results |
|
90 Day
Specifications |
The
standard specification interval for the 743B is 1 or 2 years. Typical
90 day measurement and source accuracy can be estimated by dividing
the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of f.s.", "counts" or "ohms",
remain constant. |
|