Electrical Characteristics |
|
Nominal current in: |
|
Overload: |
|
Conductor position influence: |
|
Error due to adjacent conductor: |
|
Phase shift at 2kHz at 200A: |
|
Frequency range: |
|
Temperature coefficient: |
|
Working voltage (see safety section): |
|
Order reference:
|
|
|
|
|
200Amps
AC rms |
|
up to
300Amps rms |
|
<0.5%
of range at 50/60 Hz |
|
less
than or equal to 15mA/A at 50Hz |
|
< ±1
degree |
|
40 Hz
to 10 kHz (-3 dB) |
|
0.015%
of range /°C |
|
600
Volts AC rms or DC |
|
single-phase: EP0443A
three-phase: EP0455A
three-phase+N: EP0456A |
|
|
Accuracy |
|
Primary current 0.5A to 10A:
|
|
Primary current 10A to 50A:
|
|
Primary current 50A to 100A:
|
|
Primary current 100A to 200A:
|
|
|
|
|
Operating error in % of reading: ±2.5%
Phase Shift: not specified |
|
Operating error in % of reading: ±1.5%
Phase Shift: <2° |
|
Operating error in % of reading: ±0.5%
Phase Shift: <0.5° |
|
Operating error in % of reading: ±0.5%
Phase Shift: <1° |
|
|
General Characteristics |
|
Maximum conductor size: |
|
Cable length: |
|
Operating temperature range: |
|
Storage temperature range: |
|
Operating humidity: |
|
Weight per clamp: |
|
|
|
|
15mm
diameter, bus bar 15mm x 17mm |
|
2
meters |
|
-10°C
to +55°C |
|
-20°C
to +70°C |
|
15% to
85% R.H. (non condensing) |
|
175 gms |
|
|
Reference Conditions |
|
Environmental temperature range: |
|
Humidity: |
|
Current:
|
|
Load
impedance for voltage output: |
|
|
|
|
+18°C
to +26°C |
|
20% to
75% R.H. |
|
sinusoidal waveform, 48 to 65Hz, distortion factor:
<1%, no DC component |
|
greater
than or equal to 1m ohm, <100pF |
|
|
Safety
Standards |
IEC/EN61010-1: 2001
IEC/EN61010-2-032
IEC/EN61010-2-031 |
|
Safety |
600V CAT III,
pollution degree 2
The application of the clamps on uninsulated conductors is
limited to 600V AC rms or DC to ground and frequencies below
1kHz |
|
EMC
Standards |
EN61326-1:
1997/A1: 1998 |
|
Dimensions |
102mm x
42.5mm x 23.5mm |