Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

   
Fluke/Hart Scientific Logo

1560 Black Stack Thermometer Readout
Specifications

 

The Black Stack is the perfect foundation to build a totally automated calibration system.

Model 1560 Base Unit
Power: 100 to 240 VAC, 50 or 60 Hz, nominal; Attachable Modules: up to 8; Display: 4.25" x 2.25" LCD graphics, LED backlight, adjustable contrast and brightness; Automatic Input Sequencing: 1 to 96 channels; Communications: RS-232; Non-volatile Memory: channel sequence, probe coefficients; Minimum Sample Time: 2 seconds.


 

 
Resistance Modules Specifications
SPRT Module 2560
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
2 0Ω to 400Ω ±20ppm of reading
(0.0005Ω at 25Ω,
0.002Ω at 100Ω)
0.0001Ω -260°C to 962°C ±0.005°C at 0°C
±0.007°C at 100°C
0.0001°C 1.0mA
1.4mA
 
 
High-Temp PRT Module 2561
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
2 0Ω to 25Ω ±50ppm of reading
(0.00013Ω at 2.5Ω)
0.00001Ω 0°C to 1200°C ±0.013°C at 0°C
±0.018°C at 100°C
0.001°C 3.0mA
5.0mA
 
 
PRT Scanner 2562
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
8 0Ω to 400Ω ±40ppm of reading
(0.004Ω at 100Ω)
0.0001Ω -200°C to 850°C ±0.01°C at 0°C
±0.014°C at 100°C
0.0001°C 1.0mA
1.4mA
 
 
Standards Thermistor Module 2563
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
2 0Ω to 1MΩ ±50ppm of reading
(0.5Ω at 10kΩ)
0.1Ω -60°C to 260°C ±0.0013°C at 0°C
±0.0015°C at 75°C
0.0001°C 2µA
10µA
 
 
Thermistor Scanner 2564
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
8 0Ω to 1MΩ ±100ppm of reading
(1Ω at 10kΩ)
0.1Ω -60°C to 260°C ±0.0025°C at 0°C
±0.003°C at 75°C
0.0001°C 2µA
10µA
 
 
1000 Ohm PRT Module 2567
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
2 0Ω to 4kΩ ±25ppm of reading
(0.025Ω at 1kΩ)
0.001Ω -260°C to 962°C ±0.006°C at 0°C
±0.009°C at 100°C
0.0001°C 1.0mA
1.4mA
 
 
1000 Ohm PRT Module 2568
Input
Channels
Resistance
Range
Basic
Resistance Accuracy
Resistance
Resolution
Temperature
 Range
Equivalent
Temperature
Accuracy Note 1
Temperature
Resolution
Excitation
Current
8 0Ω to 4kΩ ±40ppm of reading
(0.04Ω at 1kΩ)
0.001Ω -200°C to 850°C ±0.01°C at 0°C
±0.014°C at 100°C
0.0001°C 0.1mA
0.05mA
 
 
Thermocouple Modules
Precision Thermocouple Module 2565
Input
Channels
Millivolt
Range
Millivolt
 Accuracy
Millivolt
Resolution
Temperature
 Accuracy Note 1
Ext. CJC
Temperature
Accuracy Note 1
Int. CJC
Temperature
Resolution
2 -10mV to 100mV ±0.002mV 0.0001mV ±0.05°C ±0.1°C 0.001°C
 
 
Precision Thermocouple Module 2565
Input
Channels
Millivolt
Range
Millivolt
 Accuracy
Millivolt
Resolution
Temperature
 Accuracy Note 1
Ext. CJC
Temperature
Accuracy Note 1
Int. CJC
Temperature
Resolution
12 -10mV to 100mV ±0.004mV 0.0001mV ±0.1°C ±0.3°C 0.001°C
 
Note1: Temperature accuracy depends on probe type and temperature

 