Models 2100 and 2200 Benchtop Temperature Controller - Specifications

 

Specifications

2100 2200

Temperature Range

-100°C to 670°C -100°C to 800°C

Control Stability

±0.0005°C to ±0.002°C
(depends on system design)
±0.005°C to ±0.02°C
(depends on system design)

Display Accuracy (with probes as below) Note 1.

±1.0°C ±1.0°C

Display Resolution

0.01° 0.01°

Set-Point Resolution

0.0002° in high-resolution mode 0.01°

Auxiliary and Heater Output

100V to 125V ac nominal
(internally switchable),
50/60Hz, 10A maximum
100 to 230V ac,
50/60Hz, 10A maximum

Heater Output

Solid-state relay Solid-state relay

Dimensions

72mm (H) x 172mm (W) x 250mm (D) 72mm (H) x 114mm (W) x 178mm (D)

Automation Software

Both models include Hart's 9930 Interface-it software package

 
Note 1.
 
Performance is dependent on system design including the control sensor. Capabilities are based on factory observed performance
 

Probes

2620 2622 2624 2611Note 2. 5635 Note3.

Probe Type

RTD RTD RTD Thermistor 'K' Thermocouple

Measuring Range

-100°C to 550°C -100°C to 550°C -100°C to 550°C -10°C to 110°C 1100°C

Dimensions (mm)

280mm x 4.8mm 229mm x 4.8mm 356mm x 4.8mm 229mm x 5.5mm 406mm x 4.7mm

Dimensions (ins)

11" x 0.187" 9" x 0.187" 14" x 0.187" 9" x 0.218" 16" x 0.187
 
Note 2. Reference Probe 2611 ( use with 2100 controller only)
Note 3. Reference Probe 5635 ( used for cutout)

 