Technical Specifications |
|
Temperature Range |
35°C to 400°C
(95°F to 752°F) |
|
Display Accuracy |
±0.5°C to 200°C,
±1.0°C to 400°C |
|
Stability |
±0.2°C to 300°C,
±0.3°C to 400°C |
|
Resolution |
0.01° |
|
Uniformity |
±0.3°C at 100°C
±0.6°C at 200°C
±0.9°C at 300°C
±1.4°C at 400°C |
|
Heating Time |
25°C to 400°C: 22
minutes |
|
Cooling Time |
400°C to 100°C:
65 minutes |
|
Stabilization Time |
8 minutes |
|
Controller |
Hart model
2200. Microprocessor based, with RS-232 communication |
|
Readout |
°C or °F,
switchable |
|
Sensor |
RTD, 100 ohm |
|
Heater |
325-watt, solid-state controlled |
|
Surface Plate |
6061 aluminum; top surface machine
finished to 0.0008mm (0.000032 inch),
96mm (3.8 in) diameter accessible |
|
Power Supply |
115Vac (±10%)
50/60Hz, 2.8A or 230Vac (±10%), 50/60Hz, 1.4A, 325W
Specify voltage and frequency |
|
Weight |
3.2kg (7lb) with 2200 Controller |
|
NIST-Traceable Calibration
Data At: |
50°C, 120°C,
190°C, 260°C, 330°C, and 400°C |