Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

Fluke/Hart Scientific Logo

Precision Infrared Calibrator
Models 4180 and 4181 - Specifications

 

Model

4180 4181

Temperature Range at 23°C ambient, 0.95 emissivity

-15°C to 120°C

35°C to 500°C

Display Accuracy Note1

±0.40°C at -15°C
±0.40°C at 0°C
±0.50°C at 50°C
±0.50°C at 100°C
±0.55°C at 120°C

±0.35°C at 35°C
±0.50°C at 100°C
±0.70°C at 200°C
±1.20°C at 350°C
±1.60°C at 500°C

Stability

±0.10°C at -15°C
±0.05°C at 0°C
±0.10°C at 120°C

±0.05°C at 35°C
±0.20°C at 200°C
±0.40°C at 500°C

Uniformity Note2
(5.0" dia of center of target)

±0.15°C at -15°C
±0.10°C at 0°C
±0.20°C at 120°C

±0.10°C at 35°C
±0.50°C at 200°C
±1.00°C at 500°C

Uniformity Note2
(2.0" dia of center of target)

±0.10°C at -15°C
±0.10°C at 0°C
±0.20°C at 120°C

±0.10°C at 35°C
±0.25°C at 200°C
±0.50°C at 500°C

Heating Time

15 minutes: -15°C to 120°C
14 minutes: 23°C to 120°C

20 minutes: 35°C to 500°C
 

Cooling Time

15 minutes: 120°C to 23°C
20 minutes: 23°C to -15°C

100 minutes: 500°C to 35°C
40 minutes: 500°C to 100°C

Stabilization Time

10 minutes 10 minutes

Nominal Emissivity Note3

0.95 0.95

Thermometer Emissivity
Compensation

0.9 to 1.0

Target Diameter

152.4mm (6 inches)

Computer Interface

RS-232 interface with 9930 Interface-it software and User Guide.

Power Supply

115Vac (±10%), 6.3A,
50/60Hz, 630W
230Vac
(±10%), 3.15A,
50/60Hz, 630W
115Vac (±10%), 10A,
50/60Hz, 1000W
230Vac
(±10%), 5A,
50/60Hz, 1000W

Fuses

115Vac 6.3A, 250V slow blow
230Vac 3.15A, 250V, T
115Vac 10A, 250V fast blow
230Vac 5A, 250V, F

Dimensions (H x W x D)

356mm x 241mm x 216mm
(14" x 9.5" x 8.5")
356mm x 241mm x 216mm
(14" x 9.5" x 8.5")

Weight

9.1kg (20lb) 9.5kg (21lb)

Safety

EN 61010-1:2001, CAN/CSA C22.2 No. 61010-04
   

Note1:

For 8µm to 14µm spectral band thermometers with emissivity set between 0.9 and 1.0.

Note2:

The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the target will measure the same temperature.

Note3:

The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize emissivity related uncertainties.

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |
 Site Map | Search | About Us | Contact Us

 

Rapid-Tech Equipment
ABN 79 883 184 102
1300 Ferntree Gully Road. Scoresby. Victoria 3179
P.O. Box 4243 Knox City Centre. Victoria 3152

Phone: +61 3 9763 3299     Fax: +61 3 9753 2132