Models 5606 and 5607 Full Immersion PRTs for Laboratory, Freezers, Autoclaves and Furnaces

 

Specifications

5606

5607

Temperature Range

-200°C to 160°C

-200°C to 450°C

Nominal Resistance at 0.01°C

100Ω ±0.1Ω

100Ω ±0.1Ω

Temperature Coefficient

0.00385Ω/Ω/°C

0.00385Ω/Ω/°C

Accuracy Note1
 

±0.04°C at 0.01°C
±0.06°C at 160°C

±0.04°C at 0.01°C
±0.07°C at 450°C

Short-Term Repeatability Note2
Accuracy

±0.03°C at 0.01°C
±0.04°C at 160°C

±0.020°C at 0.01°C
±0.035°C at 450°C

Drift Accuracy Note3
 

±0.03°C at 0.01°C
±0.04°C at 160°C

±0.02°C at 0.01°C
±0.04°C at 450°C

Hysteresis

±0.015°C

Sheath Length

50mm ±5mm

40mm ±5mm

Sheath Diameter

31mm ±0.1mm

Sheath Material

316 SST

Inconel® 600

Transition Junction Temperature Range Note4

-200°C to 160°C
 

0°C to 450°C
 

Transition Junction Dimensions

No Transition

35mm x 7mm

Sensor Length

30mm ±3mm

Sensor Location

3mm ±1mm from tip

Minimum Insulation Resistance

20MΩ at 23°C

100MΩ at 23°C

Minimum Immersion Length

Full Immersion

Full Immersion

Maximum Immersion Depth in Liquid Medium

Full Immersion
 

Use SST or glass tube to protect transition junction

Maximum Immersion in
Dry Medium

Full Immersion

Full Immersion

Response Time Note5

12 seconds typical

9 seconds typical

Self Heating (in 0°C Bath)

±0.003°C

Lead-wire Cable Type

Enameled copper wire

SST over-braid

Lead-wire Length

2.4 meters

Lead-wire Temperature Range

160°C

450°C

Calibration

Calibration not included. NVLAP accredited calibration available. See ordering code.

 

Note1:

Includes calibration and 100 hour drift (k=2).

Note2:

Three thermal cycles from minimum to maximum temperature, includes hysteresis, 95% confidence (k=2).

Note3:

After 100 hours at maximum temperature, 95% confidence (k=2).

Note4:

Temperature outside this range will cause irreparable damage.

Note5:

Per ASTM E 644
 
Calibration Uncertainty (Model 5606)
Temperature
 
1924 Calibration
Uncertainty Note*
Tolerance without
Calibration**
-197°C 0.03°C 0.4°C
-38°C 0.03°C 0.2°C
0°C 0.03°C 0.1°C
100°C 0.045°C 0.3°C
157°C 0.05°C 0.4°C
 
 
Calibration Uncertainty (Model 5607)
Temperature
 
1924 Calibration
Uncertainty Note*
Tolerance without
Calibration**
0°C 0.03°C 0.1°C
157°C 0.05°C 0.2°C
232°C 0.05°C 0.3°C
450°C 0.055°C 0.5°C
 

Note*

Lab code 200348-0

Note**

Each PRT includes a certificate with a measured R0 value.

 

 

An Uncalibrated 5606 or 5607 meets ASTM E 1137 Grade A Classification of Tolerances.
Use the coefficients below to achieve the tolerances listed in the "No Calibration" column or calculate the Grade A tolerance using the following formula: Grade A = ±[0.13 + 0.0017|t|]°C
where: |t|=value of temperature (°C) without regard to sign. See ASTM E 1137.
 

R0:

Enter the R0 value provided

A:

3.9083 x 10-3

B:

-5.775 x 10-7

C:

-4.183 x 10-4

 