Specifications

5608

5609

Temperature Range

-200°C to 500°C

-200°C to 670°C

Nominal Resistance

100Ω ±0.5Ω

100Ω ±0.5Ω

Temperature Coefficient

0.0039250Ω/Ω/°C

0.0039250Ω/Ω/°C

Accuracy (Note1)

 

Short-Term Repeatability (Note2)

 

 

 

Drift (Note3)

 

 

 

Hysteresis

See footnote

±0.01°C at 0.010°C

±0.02°C at max. temperature

±0.01°C at 0.010°C

±0.02°C at max. temperature

±0.01°C maximum

See footnote

±0.01°C at 0.010°C

±0.02°C at max. temperature

±0.01°C at 0.010°C

±0.02°C at max. temperature

±0.01°C maximum

Sensor Length

30mm ±5mm (1.2" ±0.2")

Sensor Location

3mm ±1mm from tip (0.1" ±0.1")

Sheath Material

InconelTM 600

Minimum Insulation Resistance

500MΩ at 23°C

500MΩ at 23°C

 

20MΩ at 500°C

10MΩ at 670°C

Transition Junction Dimensions

71mm x 12.5mm (2.8" x 0.49")

Minimum Insertion Length (Note4)
(<5mK error)

80mm (3.1")
 

100mm 3.9")
 

Maximum Immersion Length

305mm (12")

Response Time

9 seconds typical

12 seconds typical

Self Heating (in 0°C bath)

75mW/°C

50mW/°C

Lead-Wire Cable Type

TeflonTM 24AWG

Lead-Wire Length

1.8 meters (6ft)

Lead-Wire Temperature Range

-50°C to 250°C

Calibration (optional)

 

NVLAP accredited calibration optional, Lab code 200348-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.

 

Note1

 

Accuracy is a difficult term when used to describe a resistance thermometer. The simplest way to derive basic "accuracy" is to combine the probe drift specification and calibration uncertainty with readout accuracy at a given temperature.

Note2

Three thermal cycles from minimum to maximum temperature, includes hysteresis, 99.9% confidence.

Note3

After 100 hours at maximum temperature, 99.9% confidence.

Note4

Per ASTM E 644
 

 

Calibration Uncertainty for Optional Calibrations

Temperature

1922 1923 1924

-179°C

0.010°C

0.025°C

0.025°C

-38°C

0.009°C

0.025°C

0.025°C

0°C

0.009°C

0.025°C

0.025°C

157°C

0.014°C

0.030°C

0.045°C

232°C

0.016°C

0.030°C

0.045°C

420°C

0.025°C

0.035°C

0.045°C

660°C

n/a

0.050°C

0.050°C

 

Note:






 

Uncertainties depend on lab practices, available information and equipment. A selection of calibrations are available to meet customer needs.

1930 calibrations are for complete thermometer systems.
1923 and 1924 calibrations are for PRTs only. (as of 2008, model1923 and 1930 calibrations are accredited)
1924 calibrations have not yet been accredited.
Lab code 200348-0
1922-4-R
uncertainties are ±0.025°C at 500°C

 