The 5608 comes with a one-eighth inch (3.18 mm)
diameter sheath in lengths of 9 inches or 12 inches. The 5609
comes with a one-quarter inch (6.35 mm) diameter sheath in lengths
of 12 inches, 15 inches, and 20 inches; or with a 6 mm diameter in
lengths of 300 mm, 400 mm, or 500 mm.
When looking for improved response time and
reduced stem effect in shallow immersion, look for small diameter
probes, because the measurement error called stem effect is caused
by the diameter of the stem rather than the length of the stem.
Both of these probes have Inconel™ sheaths and are
made using a special manufacturing process, giving them great
precision over a wide temperature range. The sensors for these
probes are reference-grade platinum and feature four-wire
connections with less noisy measurements than two-wire
counterparts.
These probes come with a certificate of compliance
to ensure the meet their specifications. If you’d like, you can
also order a NVLAP-accredited calibration from our laboratory, lab
code 200348-0. On the report of calibration, you’ll get the test
data and the ITS90 calibration coefficients that you can easily
input into any Hart thermometer. If you order your probe with an
INFO-CON connector, we’ll program the coefficients directly into
your connector, which loads the coefficients for you when you plug
it into our 1522 Handheld Thermometer. |