Secondary Reference Temperature Standard, PRT, Model 5615 - Specifications

 

Specifications

5615-6

5615-9

5615-12

Temperature Range

-200°C to 200°C -200°C to 420°C -200°C to 420°C

Nominal Resistance

100Ω ±0.10Ω

Temperature Coefficient

0.0039250Ω/Ω/°C

Accuracy (Note1)

 

Short-Term Repeatability (Note2)

 

 

 

Drift (Note3)

See footnote

±0.013°C at 0.010°C

 

±0.01°C at 0.010°C

Sensor Length

28mm

Sensor Location

6.9mm ±3.3mm from tip

Sheath Diameter Tolerance

±0.127mm

Sheath Material

InconelTM 600

Minimum Insulation Resistance

1000MΩ at 23°C

Transition Junction Temp Range (Note4)

-50°C to 200°C

Transition Junction Dimensions

71mm x 13mm

Maximum Immersion Length

102mm 178mm 254mm

Response Time (Note5)

9 seconds typical

Self Heating (in 0°C bath)

50mW/°C

Lead-Wire Cable Type

TeflonTM insulated with TeflonTM jacket, 22 AWG

Lead-Wire Length

183cm

Lead-Wire Temperature Range

-50°C to 200°C

Calibration (optional)

 

NVLAP accredited calibration included, Lab code 200706-0. Please see calibration uncertainty table and its explanation of changeable uncertainties.

 

Note1

 

Accuracy is a difficult term when used to describe a resistance thermometer. The simplest way to derive "accuracy" is to combine the probe drift specification and calibration uncertainty with readout accuracy at a given temperature.

Note2

Three thermal cycles from minimum to maximum temperature, includes hysteresis, 99.9% confidence.

Note3

After 100 hours at maximum temperature, 99.9% confidence.

Note4
 

Temperatures outside this range will cause irreparable damage. For best performance, transition junction should not be too hot to touch.

Note5

Per ASTM E 644
 

Footnote (NVLAP*) Calibration Uncertainty

Temperature Expanded Uncertainty (K = 2)  

-196°C

0.024°C

Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties.

-38°C

0.011°C

0°C

0.010°C

200°C

0.018°C

420°C

0.029°C

 

NVLAP*

Lab code 200706-0

5615-6

excluded

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |
 Site Map | Search | About Us | Contact Us

 

Rapid-Tech Equipment
ABN 79 883 184 102
1300 Ferntree Gully Road. Scoresby. Victoria 3179
P.O. Box 4243 Knox City Centre. Victoria 3152

Phone: +61 3 9763 3299     Fax: +61 3 9753 2132