Model 5623B Precision RTD Freezer Probe
Specifications

 

Parameter

Value

Temperature Range

-100°C to 156°C

Transition Temperature

-100°C to 156°C

Nominal Resistance at 0°C

100Ω (±0.1Ω)

Temperature Coefficient

0.003925Ω/Ω/°C nominal

Drift
 

±0.01°C per year maximum at 0°C, when used periodically at maximum temperature.

Sheath Material

InconelTM 600

Leads

22 AWG TeflonTM-insulated, silver plated stranded copper

Termination

Specify- see Ordering Information

Calibration
 

Includes manufacturer's NVLAP-accredited, lab code 200706-0 calibration and table with Rvs. T values in 1°C increments from -80°C to 156°C.
ITS-90 coefficients included.

Dimensions

152mm x 6.35mm

Cable Length

6.1 meters

Calibrated Accuracy (k=2)

±0.05°C over the full range.  Includes calibration uncertainty and drift.

 