Ideal for use as a reference thermometer in high-temperature
furnaces, the 5624 can reach a temperature of 1000°C with
long-term drift at 0°C of 10 mK and accuracy (including
short-term stability and calibration uncertainty over the
full range) of 55 mK! Due to Hart Scientifics' proprietary
sensor design, this PRT has short-term stability of 5 mK,
and an immersion requirement of less than 153 mm (6") at
700°C.
The 5624 is assembled in an alumina sheath that is 508 mm
(20") long and 6.35 mm (0.25") in diameter. Several
termination configurations can be selected to match
different thermometer readouts. Each 5624 comes with a
NIST-traceable, NVLAP-accredited fixed-point calibration
from 0°C to 962°C. That’s a fixed-point calibration! The
5624 also comes in a protective carrying case. |