The 5640, 5641, and 5642 thermistor probes are designed for
the temperature range of 0°C to 60°C. The 5643 and 5644
probes span the 0°C to 100°C temperature range. They offer
stability of either ±0.002°C or ±0.005°C. These stability
levels are guaranteed for one full year.
Precision calibration, traceable to NIST, is provided with
each probe. A computer-generated table in increments of
0.01°C is furnished with each calibration based on the
formula:
R=exp(A + B/T + C/T2 + D/T3)
The constants for the formula are obtained from a polynomial
regression performed on the calibration data obtained. Over
the range of 0°C to 60°C, calibration is performed at the
triple point of water (0.01°C) and 15°C, 25°C, 30°C,
37°C, 50°C and 60°C. For the 0°C to 100°C temperature range,
the additional calibration points of 80°C and 100°C are
used.
Each probe is individually calibrated and includes a report
of calibration from the manufacturer. Contact Hart for
calibration in Hart’s NVLAP accredited lab.
Thermistor standards are rugged, precision sensors suitable
for use as secondary or working temperature standards for
laboratory metrology applications. Because they generally
are not affected by shock and vibration, you can use them in
the most difficult field environments without worrying about
calibration integrity.
Combine these probes with Hart’s 1560 Black Stack
thermometer to read directly in °C, °F, or K. This
combination gives you resolution of 0.0001 degrees and total
system accuracy is better than ±0.004°C.
Compare the cost of a 5640 calibrated probe and a Black
Stack thermometer to the cost of one Uncalibrated SPRT.
Between 0°C and 100°C, nothing beats the value of the 5640
Series Thermistors. |