5698 Working Standard SPRT
-200°C to 670°C Thermometer

 

SPRTs. Art or science? It takes, in fact, quite a bit of both. The one thing not involved is mystery. That's why HART SPRTs always include detailed published specifications, including drift rates. Hart Scientific newest SPRT is no exception. It was designed by the same Hart metrologists who have created a dozen different SPRT designs used in national labs around the world. And it performs just like they say it does.

Model 5698 Working Standard SPRT

 

Features

  • Temperatures from -200°C to 670°C.
  • Fully conforms to ITS-90 SPRT guidelines.
  • Drift rate typically less than 0.003°C.
  • Multiple calibration options by fixed point.
  • Unmatched performance-to-price-ratio

 

 

 

Description
The 5698 Working Standard SPRT is a true SPRT. It meets the ITS-90 ratio requirements for SPRTs and includes a Hart designed, completely strain-free platinum sensor. With 485-mm quartz sheath, this 25-ohm SPRT covers a temperature range from -200°C to 670°C. Long-term drift, which we define as the change in output resistance at the triple point of water after 100 hours at 670°C, is (after converting to temperature) less than 6mK-typically less than 3mK.

The 5698 is the perfect companion to a Hart Super-Thermometer such as the Model 1590, which reads 25-ohm SPRTs to within 1mK at 0°C and includes a number of convenient features for working with SPRTs. Requiring 1mA of excitation current, the 5698 can also be used easily with a Hart Black Stack, or even a Chub-E4 Thermometer.

At Hart Scientific, they provide excellent post-purchase support so you have the best possible ownership experience. They also provide all the information before you purchase including detailed performance specifications.

Maybe there is some art mixed with their science. But that doesn't mean they keep secrets. Trust your Lab Standards to Hart Scientific.

 

Specification

Description

Temperature Range

-200°C to 670°C

Nominal RTPW

25.5Ω (±0.5Ω)

Current

1.0mA

Resistance Ratio

W(234.315k)≤0.844235
W(302.914k)≥1.11807

Sensitivity

0.1Ω/°C

Drift Rate

<0.006°C/100 hours at maximum temperature (typically <0.003°C)

Self Heating at TPW

<0.002°C under 1mA current

Reproducibility

±0.0015°C or better

RTPW Drift after Thermal Cycling

<0.001°C

Lead Wires

Four sensor wires (Diameter of Pt sensor wire 0.07mm)

Protective Sheath

Quartz Glass 7mm diameter x 485mm long

 
 

Ordering Information

Hart 5698-25

25Ω Working Standard SPRT, -200°C to 670°C, with Maple carrying case.

 