Specifications |
7077 |
6054 |
6055 |
|
Range |
-5°C to
110°C |
50°C to
300°C |
200°C to
550°C |
|
Stability |
±0.001°C at 0°C
(ethanol)
±0.003°C at 100°C
(oil 5012) |
±0.003°C at 100°C
(oil 5012)
±0.005°C at 300°C
(oil 5017) |
±0.003°C at 200°C
(salt)
±0.01°C at 550°C
(salt) |
|
Uniformity |
±0.004°C at 0°C
(ethanol)
±0.007°C at 100°C
(oil 5012) |
±0.007°C at 100°C
(oil 5012)
±0.015°C at 300°C
(oil 5017) |
±0.005°C at 200°C
(salt)
±0.010°C at 550°C
(salt) |
|
Temperature Setting |
Digital display
with push-button data entry |
|
Set-Point Resolution |
0.002°C,
high resolution 0.00003°C |
0.01°C,
high resolution 0.00018°C |
|
Display Temperature
Resolution |
0.01°C |
|
Digital Setting Accuracy |
±1°C |
|
Digital Setting Repeatability |
±0.005 °C |
±0.01°C |
|
Heaters |
250 watts to 1000 watts |
250 watts to 1000 watts |
225 watts to 1800 watts |
|
Working Area |
178mm dia. (7 in) |
196mm dia. (7.7 in) |
107mm dia. (4.2 in) |
|
Depth
|
610 mm (24 in) deep,
178 mm dia.
(7 in), removable polycarbonate cover |
610 mm deep (24 in),
196 mm dia. (7.7 in), removable SST lid |
432 mm deep (17 in),
107 mm dia. (4.2 in), removable SST lid, special viewing
channel for LIG sighting |
|
Wetted Parts |
304 Stainless
Steel |
|
Power Supply |
230 VAC (±10 %),
50Hz or 60 Hz, 14 A
(Specify frequency, contact Hart if CE mark required.) |
230 VAC (±10 %),
50/60 Hz, 10.7 A |
230 VAC (±10 %),
50/60 Hz, 7.8 A |
|
Volume |
42 liters |
50 liters |
19.8 liters
43kg of Bath Salt |
|
Dimensions (D x W x H) |
470 x 775 x 194 mm to working
surface (18.5 x 30.5 x 47 in),
1397 mm (55 in) to top of stir motor, 914 mm (36 in) to
control panel |
572 x 762 x 1219 mm to
working surface (22.5 x 30 x 48 in), 1422 mm (56 in) to top
of stir motor box, 914 mm (36 in) to control panel |
572 x 775 x 1219 mm to
working surface (22.5 x 30.5 x 48 in), 1524 mm (60 in) to
top of stir motor box, 914 mm (36 in) to control panel |
|
Distance from Line of Sight
to Top of Fluid |
9.3mm |
15.9mm |
15.9mm |
|
Automation Package |
Interface-it
software and RS-232 computer interface is optional for
setting bath temperature via remote computer. For IEEE-488,
add the 2001-IEEE to the automation package |