Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

   
Fluke/Hart Scientific Logo

Models 7009, 7015, and 7108
Resistor Bath - Specifications

 

Specifications

7015 7009 7108

Range

0°C to 50°C 0°C to 50°C 20°C to 30°C

Stability at 25°C

±0.0007°C (water)
 ±0.001°C
(mineral oil 5011)
±0.0007°C (water)
 ±0.001°C
(mineral oil 5011)
±0.002°C (water)
 ±0.004°C
(mineral oil 5011)

Uniformity at 25°C

±0.003°C (water)
 ±0.005°C
(mineral oil 5011)
±0.003°C (water)
 ±0.005°C
(mineral oil 5011)
±0.005°C (water)
 ±0.008°C
(mineral oil 5011)

Temperature Setting

Digital display with push-button data entry

Set-Point Resolution

0.001°C, high resolution 0.00003°C 

Display Temperature Resolution

0.01°C

Display Setting Accuracy

±1°C ±1°C ±0.5°C

Digital Setting Repeatability

±0.01°C

Heaters

500 and 1000 watts 500 and 1000 watts Peltier heating/cooling

Cooling Capacity

100 to 200 watts 100 to 200 watts 100 watts in ambient
23°C

Access Opening

699mm x 279mm
(27.5" x 11")
699mm x 559mm
(27.5" x 22")
356mm x 356mm
(14" x 14")

Bath Chamber
Dimensions (H x W x D)
(unobstructed space)
 

699mm x 279mm x 330mm
(27.5" x 11" x 13")
699mm x 599mm x 330mm
(27.5" x 22" x 13")
356mm x 203mm x 356mm
(14" x 8" x 14")

Depth

330mm (13") 330mm (13") 203mm (8")

Wetted Parts

304 Stainless Steel 304 Stainless Steel 304 Stainless Steel
Resistor rack: Hard-anodized, perforated aluminum

Safety Cutout

Factory-set
high temperature
Factory-set
high temperature
n/a

Power Supply

115V AC (±10 %)
60Hz 15A
or
 230 VAC (±10 %),
50Hz or 60 Hz, 14 A
(Specify) 
230 VAC (±10 %),
50Hz or 60 Hz, 12A
(specify frequency)
115V AC (±10 %)
50/60Hz 3A
or
 230 VAC (±10 %),
50/60 Hz, 1.6 A
(Specify) 

Volume

95 liters (25 gallons)

167 liters (44 gallons) 51 liters (13.2 gallons)

Dimensions (H x W x D)

1219 x 1118 x 559mm 1092 x 1130 x 864mm 489 x 413 x 635mm

Weight

141kg 150kg 35kg

Automation Package

Interface-it software and RS-232 computer interface are available for setting bath temperature via external computer. (both come standard with model 7108). For IEEE-488, add the 2001-IEEE to the automation package
 

Note: Although the 7015 and 7009 baths are capable of reaching higher temperatures, they are designed for use with Standard Resistors. Therefore, the soft cutout of the instrument has been set at the factory to 50°C to protect standard resistors placed in the bath.

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |
 Site Map | Search | About Us | Contact Us

 

Rapid-Tech Equipment
ABN 79 883 184 102
1300 Ferntree Gully Road. Scoresby. Victoria 3179
P.O. Box 4243 Knox City Centre. Victoria 3152

Phone: +61 3 9763 3299     Fax: +61 3 9753 2132