Features and Benefits
The Model 7900 includes a special circuit that monitors the controller’s
microprocessor and automatically resets it if its operations are interrupted.
Two separate cutout systems are also included for keeping your bath’s
temperature within its normal range.
A software
cutout uses an adjustable high-temperature limit that can be easily accessed
through the front panel and set to match the requirements of your bath fluid.
Should the control sensor measure a temperature beyond this upper limit,
heating is shut down. If the bath’s temperature falls below its normal
operating range, the heaters are turned on and the LN
2 cooling is shut off. A second, independent hardware cutout
monitors the bath’s temperature with a thermocouple and shuts down all heating
and LN
2 cooling if the bath’s temperature rises above its
range.
These cutout
features, combined with the superior reliability and long-term stability
performance of the 7900, allow you to run your system for as long as you like
between shut-downs—365 days a year, if you wish. Your bath can be ready for
you to take measurements the minute you walk into your lab each day. |