Specifications |
Hot Block |
Cold Block |
|
Range |
50°C to 350°C
(122°F to 662°F) |
-15°C to 110°C
(15°F to 230°F)
-8°C with hot block at 350°C
18°F with hot block at 662°F |
|
Accuracy |
±0.6°C |
±0.2°C |
|
Stability |
±0.1°C |
|
Well-to-Well Uniformity |
±0.1°C |
|
Display Resolution |
±0.1° |
|
Heating Times |
30 minutes from
25°C to 350°C |
15 minutes from
25°C to 110°C |
|
Cooling Times |
40 minutes from
350°C to 100°C |
16 minutes from
25°C to -15°C |
|
Stabilization Times |
8 minutes |
|
Well Depth |
102mm (4") |
|
Removable Inserts |
Two 6.4mm (1/4") and
two 4.8mm (3/16") inserts included;
see Ordering Information for other available inserts |
|
Power Supply |
115Vac (±10%),
3A
or
230Vac (±10%), 2A, 280W
Also specify 50/60Hz and Voltage |
|
Dimensions |
178mm (H) x 267mm (W) x
248mm (D)
7" (H) x 10.5" (W) x 9.75" (D) |
|
Weight |
4.5kg (10lb) |
|
Computer Interface |
RS-232 Interface
included with free Interface-it software |
|
NIST-Traceable
Calibration Data at: |
50°C, 100°C,
150°C, 200°C, 250°C,
300°C and 350°C |
-8°C, 0°C,
25°C, 50°C, 75°C,
100°C and 110°C |