Specifications |
Hot Block |
Cold Block |
|
Temperature Range |
50°C
to 670°C
122°F to 1238°F |
-30°C
to 140°C
-22°F to 284°F |
|
Accuracy |
±0.2°C at
50°C
±0.4°C at 400°C
±0.65°C at 600°C |
±0.25°C
(insert wells)
±0.65°C (fixed wells)
|
|
Stability |
±0.02°C
at 100°C
±0.06°C at 600°C |
±0.02°C
at -30°C
±0.04°C at 140°C |
|
Uniformity |
±0.02°C (±0.05°C typical) |
±0.05°C
(insert wells)
±0.25°C (fixed wells) |
|
Well Depth |
152mm (6") |
124mm (4.875") |
|
Heating Time to Max. |
30 minutes |
15 minutes |
|
Cooling Times |
120 minutes from 660°C
to 100°C |
30 minutes from 140°C
to -30°C |
|
Well Inserts |
One interchangeable well
accommodates multi-hole insert |
One interchangeable well
accommodates multi-hole insert, plus
1/4", 3/16", and 1/8" |
|
Computer Interface |
RS-232 interface
included with Model 9930 Interface-it-control
software |
|
Power Supply |
115Vac (8.8A), or
230Vac (4.4A) switchable, (±10%)
50/60Hz, 1150W |
|
Dimensions |
292mm (H) x 394mm
(W) x 267mm (D)
(11.5" x 15.5" x 10.5") |
|
Weight |
16.4kg (36lb) |
|
NIST-Traceable Certificate
(8 points) Data At: |
50°C,
100°C, 200°C, 300°C, 400°C
500°C, 600°C, and 660°C |
-30°C,
0°C, 25°C, 50°C, 75°C
100°C, 125°C, and 140°C |