Models |
9100S |
9102S |
| Range |
35°C to 375°C
(95°F to 707°F) |
-10°C to 122°C
(14°F to 252°F) at 23°C ambient |
| Accuracy |
±0.25°C at
50°C; ±0.25°C at 100°C;
±0.5°C at 375°C |
±0.25°C |
| Stability |
±0.07°C at
50°C; ±0.1°C at 100°C;
±0.3°C at 375°C |
±0.05°C |
| Well-to-Well Uniformity |
±0.2°C with sensors of similar size at equal depths within wells |
| Heating Times |
35°C to 375°C:
9.5 minutes |
ambient to
100°C: 10 minutes |
| Stabilization |
5 minutes |
7 minutes |
| Cooling Times |
375°C to
100°C: 14 minutes |
ambient to
0°C: 10 minutes |
| Well Depth |
102mm (4"):
1.6mm hole is 89mm deep
(1/16" hole is 3.5" deep) |
102mm (4") |
| Removable Inserts |
N/A |
Available in sizes from 1.6mm to
11.1mm
(1/16" to 7/16")
[6.4mm (1/4") and 4.8mm (3/16") included] |
| Power Supply |
115Vac (±10%),
55Hz to 65Hz, 1.5A
or
230Vac (±10%), 45Hz to 55Hz, 175W |
94Vac to 234Vac ((±10%),
50/60Hz, 60W
or
12 volts DC |
| Dimensions |
67mm (H) x 125mm (W) x 150mm (D)
2.25" (H) x 4.9" (W) x 6.1" (D) |
99mm (H) x 140mm (W) x 175mm (D)
3.9" (H) x 5.5" (W) x 6.9" (D) |
| Weight |
1.2kg (2lb 3 oz) |
1.8kg (4lb) |
| Computer Interface |
RS-232 Interface
included with
free Interface-it software |
NIST-Traceable
Calibration Data At |
50°C, 100°C,
150°C, 200°C, 250°C,
300°C and 375°C |
-10°C, 0°C,
25°C, 50°C, 75°C,
100°C and 122°C |