Models |
9133 |
9132 |
|
Temperature Range |
-30°C to
150°C at 23°C ambient
(-22°F to 302°F at 73°F ambient) |
50°C to
500°C
(122°F to 932°F) |
|
Accuracy |
±0.4°C
(±0.72°F) |
±0.5°C
at 100°C (±0.9°F at 212°F)
±0.8°C at 500°C (±1.4°F at 932°F) |
|
Stability |
±0.1°C
(±0.18°F) |
±0.1°C
at 100°C (±0.18°F at 212°F)
±0.3°C at 500°C (±0.54°F at 932°F) |
|
Target Size |
57mm (2.25") |
|
Target Emissivity |
0.95 (±0.02
from 8 to 14µm) |
|
Resolution |
0.1° |
|
Heating Time |
15 minutes (25°C
to 150°C) |
30 minutes (50°C
to 500°C) |
|
Cooling Time |
15 minutes (25°C
to -20°C) |
30 minutes (500°C
to 100°C) |
|
Computer Interface |
RS-232 Interface
included with 9930 Interface-it software |
|
Power Supply |
115Vac (±10%),
50/60Hz, 1.5A
or
230Vac (±10%), 50/60Hz, 1.0A
switchable, 200W |
115Vac (±10%),
50/60Hz, 3A
or
230Vac (±10%), 50/60Hz, 1.5A
switchable, 340W |
|
Dimensions |
152mm (H) x 286mm (W) x 267mm
(D)
(6" x 11.25" x 10.5") |
102mm (H) x 152mm (W) x 178mm
(D)
(4" x 6" x 7") |
|
Weight |
4.6kg (10lb) |
1.8kg (4lb) |
|
NIST-Traceable Contact
Calibration Data at: |
-30°C, 0°C,
25°C, 75°C, 100°C,
125°C and 150°C |
50°C, 100°C,
200°C, 250°C, 300°C,
400°C and 500°C |