Field Metrology Wells - Models 9142, 9143 and 9144 - Specifications

 
Base Unit Specifications
 

Model

9142 9143 9144

Temperature
Range at 23°C

-25°C to 150°C
(-13°F to 302°F)

33°C to 350°C
(91°F to 662°F)

50°C to 660°C
(122°F to 1220°F)

Display Accuracy

±0.2°C Full Scale

±0.2°C Full Scale

±0.35°C at 50°C
±0.35°C at 420°C
±0.5°C at 660°C

Stability

±0.01°C Full Scale

±0.02°C at 33°C
±0.02°C at 200°C
±0.03°C at 350°C

±0.03°C at 50°C
±0.05°C at 420°C
±0.05°C at 660°C

Axial Uniformity
at 40mm (1.6")

±0.05°C Full Scale

±0.04°C at 33°C
±0.1°C at 200°C
±0.2°C at 350°C

±0.05°C at 50°C
±0.35°C at 420°C
±0.5°C at 660°C

Radial Uniformity

±0.01°C Full Scale

±0.01°C at 33°C
±0.015°C at 200°C
±0.02°C at 350°C

±0.02°C at 50°C
±0.05°C at 420°C
±0.10°C at 660°C

Loading Effect
(with a 6.35mm ref probe and three 6.35mm probes)

±0.006°C Full Scale

±0.015°C Full Scale

±0.015°C at 50°C
±0.025°C at 420°C
±0.035°C at 660°C

Hysteresis

0.025 0.03 0.1

Operating Conditions

0°C to 50°C, 0% to 90% RH (non-condensing)

Environmental Conditions (for all specifications except Temperature range)

13°C to 33°C

Immersion
(Well Depth)

150mm (5.9")

Insert OD

30mm (1.8") 25.3mm (!.00") 24.4mm (0.96")

Heating Time

16 minutes: 23°C to 140°C
23 minutes: 23°C to 150°C
25 minutes: -25°C to 150°C

5 minutes: 33°C to 350°C

15 minutes: 50°C to 660°C

Cooling Time

15 minutes: 23°C to -25°C
25 minutes: 150°C to -23°C

32 minutes: 350°C to 33°C
14 minutes: 350°C to 100°C

35 minutes: 660°C to 50°C
25 minutes: 660°C to 100°C

Resolution

0.01°

Display

LCD, °C or °F user-selectable

Dimensions

290mm (H) x 185mm (W) x 295mm (D)
11.4" (H) x 7.3" (W) x 11.6" (D)

Weight

8.16kg (18lb) 7.3kg (16lb) 7.7kg (17lb)

Power Requirements

100V to 115V AC
(±10%) 50/60 Hz, 635W
230V AC
(±10%) 50/60 Hz, 575W

100V to 115V AC
(±10%) 50/60 Hz, 1400W
230V AC
(±10%) 50/60 Hz, 1800W

Computer Interface

RS-232  interface and 9930 Interface-it Control software included
 
 

-P Specifications

Built-in Reference Thermometer Readout Accuracy (4-wire Reference Probe Note 1.

±0.013°C at -25°C
±0.015°C at 0°C
±0.020°C at 50°C
±0.025°C at 150°C
±0.030°C at 200°C
±0.040°C at 350°C
±0.050°C at 420°C
±0.070°C at 660°C

Reference Resistance Range

0 ohms to 400 ohms

Reference Resistance Accuracy Note 1.

0 ohms to 42 ohms:

±0.0025 ohms

42 ohms to 400 ohms:

±60ppm of reading

Reference Characterizations

ITS-90, CVD, IEC-751, Resistance

Reference Measurement Capacity

4-wire

Reference Probe Connection

6-pin DIN with Infocon Technology

Built-in RTD
Thermometer Readout Accuracy

NI-120:

±0.015°C at 0°C

PT-100 (385):

±0.02°C at 0°C

PT-100 (3926):

±0.02°C at 0°C

PT-100 (JIS):

±0.02°C at 0°C

RTD Resistance Range

0 ohms to 400 ohms

RTD Resistance
Accuracy Note 2.

0 ohms to 25 ohms:

±0.002 ohms

25 ohms to 400 ohms:

±80ppm of reading

RTD Characterizations

PT-100 (325), (JIS), (3926), NI-120, Resistance

RTD Measurement Capacity

4-wire RTD
(2-, 3-wire RTD w/Jumpers only)

RTD Connection

4 terminal input

Built-in TC
Thermocouple Readout
Accuracy

Type J:

±0.7°C at 660°C

Type K:

±0.8°C at 660°C

Type T:

±0.8°C at 400°C

Type E:

±0.7°C at 660°C

Type R:

±1.4°C at 660°C

Type S:

±1.5°C at 660°C

Type M:

±1.4°C at 660°C

Type L:

±0.7°C at 660°C

Type U:

±0.75°C at 660°C

Type N:

±0.9°C at 660°C

Type C:

±1.1°C at 660°C

TC Millivolt Range

-10mV to 75mV

Voltage Accuracy

0.025% of reading + 0.01mV

Internal Cold Junction Compensation Accuracy

±0.5°C (ambient of 13°C to 33°C

TC Connection

Small connectors

Built-in mA Readout Accuracy

0.02% of reading +2µV

mA Range

Cal 4 to 22mA, Spec 4 to 24mA

mA Connection

2 terminal input

Loop Power Function

24V DC loop power

Built-in Electronics Temperature Coefficient
(0°C to 13°C, 33°C to 50°C)

±0.005% of range per °C

   

Note 1.:

 

The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.

 

 

Note 2.:

 

Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires. 

 