Rapid-Tech Equipment Logo Combo Kit Power Quality Recorder Power Quality Analyzer Multifunction Digital Multimeter

Nationwide - Call 1800 358 531 

Fluke/Hart Scientific Logo

9170 Series Metrology Wells -
Specifications

Models

9170 9171 9172 9173

Range
(at 23°C ambient)

-45°C to 140°C
(-49°F to 284°F)
-30°C to 155°C
(-22°F to 311°F)
35°C to 425°C
(95°F to 797°F)
50°C to 700°C
(122°F to 1292°F)

Display Accuracy

±0.1°C full range

±0.1°C full range

±0.1°C at 100°C
±0.15°C at 225°C
±0.2°C at 425°C

±0.2°C at 425°C
±0.25°C at 660°C

Stability

±0.005°C full range

±0.005°C full range

±0.005°C at 100°C
±0.008°C at 225°C
±0.01°C at 425°C

±0.005°C at 100°C
±0.01°C at 425°C
±0.03°C at 700°C

Axial Uniformity
(60mm)

±0.01°C at -45°C
±0.04°C at -35°C
±0.02°C at 0°C
±0.07°C at 140°C

±0.025°C at -30°C
±0.02°C at 0°C
±0.07°C at 155°C

±0.05°C at 100°C
±0.1°C at 225°C
±0.2°C at 425°C

±0.1°C at 100°C
±0.25°C at 425°C
±0.4°C at 700°C

Radial Uniformity

±0.01°C full range

±0.01°C full range

±0.01°C at 100°C
±0.02°C at 225°C
±0.025°C at 425°C

±0.01°C at 100°C
±0.025°C at 425°C
±0.04°C at 700°C

Loading Effect (with a
6.35mm ref. probe and
three 6.35mm probes

±0.02°C at -45°C
±0.005°C at -35°C
±0.01°C at 140°C

±0.005°C at -30°C
±0.005°C at 0°C
±0.01°C at 155°C

±0.01°C full range

±0.02°C at 425°C
±0.04°C at 700°C

Hysteresis

0.025°C 0.025°C 0.04°C 0.07°C

Well Depth

160mm (6.3") 203mm (8") 203mm (8") 203mm (8")

Resolution

001°C

Display

LCD, °C or °F, user selectable

Key Pad

Ten keys with decimal and +/- button. Function keys, menu key, and °C/°F key

Cooling Time

44 min:
23°C to -45°C
19 min:
23°C to -30°C
19 min:
140°C to 23°C

30 min:
23°C to -30°C
25 min:
155°C to 23°C

220 min:
425°C to 35°C
25 min:
155°C to 23°C

235 min:
700°C to 50°C
153 min:
700°C to 100°C

Heating Time

32 min:
23°C to 140°C
45 min:
-45°C to 140°C

44 min:
23°C to 155°C
56 min:
-30°C to 155°C

27 min:
35°C to 425°C

46 min:
50°C to 700°C

Dimensions

366mm (H) x 203mm (W) x 323mm (D)
(14.4" x 8" x 12.7")

Weight

15kg (33lb) 15kg (33lb) 13.2kg (29lb) 15kg (33lb)

Power Supply
(specify volts/freq)

115Vac (±10%), 50/60Hz,
or
230Vac (±10%), 50/60 Hz, 550 W 

115Vac (±10%), 50/60Hz,
or
230Vac (±10%), 50/60 Hz, 1025 W 

Computer Interface

RS-232 Interface with 9930 Interface-it control software included

NIST Traceable Calibration Data At:

-45°C, 0°C, 50°C,
100°C, and 140°C

-30°C, 0°C, 50°C,
100°C, and 155°C

100°C, 150°C, 250°C,
350°C, and 425°C

100°C, 200°C, 350°C,
500°C, and 660°C

 

Note: Calibrated to 660°C; reference thermometer recommended at higher temperatures.

 
 

Specifications

Built-in Reference Input

Temperature Range

-200°C to 962°C  (-328°F to 1764°F)

Resistance Range

0 Ohms to 400 Ohms, auto-ranging

Characterizations

ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11 Callendar-Van Dusen (CVD): Ro, a, β, δ

Resistance Accuracy

0Ω to 20Ω: 0.0005Ω
20Ω to 400Ω: 25ppm

Temperature Accuracy
(does not include probe
uncertainty)

    10Ω PRTs: 25Ω and 100Ω PRTs    
    ±0.013°C at 0°C ±0.007°C at -100°C    
    ±0.014°C at 155°C ±0.006°C at 0°C    
    ±0.019°C at 425°C ±0.011°C at 155°C    
    ±0.028°C at 700°C ±0.013°C at 225°C    
      ±0.019°C at 425°C    
      ±0.027°C at 661°C    

Resistance Resolution

0Ω to 20Ω: 0.0001Ω
20Ω to 400Ω: 0.001Ω

Measurement Period

1 second

Probe Connection

4-wire with shield, 5-pin DIN connector

Calibration

NVLAP accredited (built-in reference input only), NIST-traceable calibration provided

 


Fluke | Hart | Reed | Yokogawa | EuroSMC | Kyoritsu | Megger | | Tenmars | Tektronix | Amprobe | CIE | Silvertronic |
 Site Map | Search | About Us | Contact Us

 

Rapid-Tech Equipment
ABN 79 883 184 102
1300 Ferntree Gully Road. Scoresby. Victoria 3179
P.O. Box 4243 Knox City Centre. Victoria 3152

Phone: +61 3 9763 3299     Fax: +61 3 9753 2132