Models |
9170 |
9171 |
9172 |
9173 |
|
Range
(at 23°C ambient) |
-45°C to
140°C
(-49°F to 284°F) |
-30°C to
155°C
(-22°F to 311°F) |
35°C to
425°C
(95°F to 797°F) |
50°C to
700°C
(122°F to 1292°F) |
|
Display Accuracy |
±0.1°C full range |
±0.1°C full range |
±0.1°C at 100°C
±0.15°C at 225°C
±0.2°C at 425°C |
±0.2°C at 425°C
±0.25°C at 660°C |
|
Stability |
±0.005°C full range |
±0.005°C full range |
±0.005°C at 100°C
±0.008°C at 225°C
±0.01°C at 425°C |
±0.005°C at 100°C
±0.01°C at 425°C
±0.03°C at 700°C |
|
Axial Uniformity
(60mm) |
±0.01°C at -45°C
±0.04°C at -35°C
±0.02°C at 0°C
±0.07°C at 140°C |
±0.025°C at -30°C
±0.02°C at 0°C
±0.07°C at 155°C |
±0.05°C at 100°C
±0.1°C at 225°C
±0.2°C at 425°C |
±0.1°C at 100°C
±0.25°C at 425°C
±0.4°C at 700°C |
|
Radial Uniformity |
±0.01°C full range |
±0.01°C full range |
±0.01°C at 100°C
±0.02°C at 225°C
±0.025°C at 425°C |
±0.01°C at 100°C
±0.025°C at 425°C
±0.04°C at 700°C |
|
Loading Effect (with a
6.35mm ref. probe and
three 6.35mm probes |
±0.02°C at -45°C
±0.005°C at -35°C
±0.01°C at 140°C |
±0.005°C at -30°C
±0.005°C at 0°C
±0.01°C at 155°C |
±0.01°C full range |
±0.02°C at 425°C
±0.04°C at 700°C |
|
Hysteresis |
0.025°C |
0.025°C |
0.04°C |
0.07°C |
|
Well Depth |
160mm (6.3") |
203mm (8") |
203mm (8") |
203mm (8") |
|
Resolution |
001°C |
|
Display |
LCD,
°C or °F, user selectable |
|
Key Pad |
Ten keys with
decimal and +/- button. Function keys, menu key, and
°C/°F key |
|
Cooling Time |
44 min:
23°C
to -45°C
19 min:
23°C
to -30°C
19 min:
140°C
to 23°C |
30 min:
23°C
to -30°C
25 min:
155°C
to 23°C |
220 min:
425°C
to 35°C
25 min:
155°C
to 23°C |
235 min:
700°C
to 50°C
153 min:
700°C
to 100°C |
|
Heating Time |
32 min:
23°C
to 140°C
45 min:
-45°C
to 140°C |
44 min:
23°C
to 155°C
56 min:
-30°C
to 155°C |
27 min:
35°C
to 425°C |
46 min:
50°C
to 700°C |
|
Dimensions |
366mm (H) x 203mm
(W) x 323mm (D)
(14.4" x 8" x 12.7") |
|
Weight |
15kg (33lb) |
15kg (33lb) |
13.2kg (29lb) |
15kg (33lb) |
|
Power Supply
(specify volts/freq) |
115Vac (±10%),
50/60Hz,
or
230Vac (±10%), 50/60 Hz, 550 W |
115Vac (±10%), 50/60Hz,
or
230Vac (±10%), 50/60 Hz, 1025 W |
|
Computer Interface |
RS-232 Interface
with 9930 Interface-it control software included |
|
NIST Traceable Calibration Data At: |
-45°C, 0°C, 50°C,
100°C, and 140°C |
-30°C, 0°C, 50°C,
100°C, and 155°C |
100°C, 150°C, 250°C,
350°C, and 425°C |
100°C, 200°C, 350°C,
500°C, and 660°C |