Hart’s 9260 Mini Fixed-Point Cell Furnace provides a
fixed-point system that cuts in half the financial investment
required to do fixed-point calibrations and virtually all the time
and training required by traditional systems.
This furnace costs less than half of a large furnace and works
with indium, tin, zinc, and aluminum cells to cover all ITS-90
fixed points from 156.5985 °C to 660.323°C. The cells themselves,
using a smaller volume of 99.9999% pure metal, also cost much
less. But cost is only a part of the issue. |

|
The
9260 makes using fixed points easy. Simply insert the cell at the
end of the day and let it sit overnight. The next morning,
initialize the built-in software routine for your specific cell.
Come back in an hour, verify the stability of the cell, and you
can take measurements for the rest of the day from a near-perfect
temperature source!
The
built-in software lets you choose between using melting-point
curves or freezing-point curves for each metal. The ITS-90 calls
for freezing points, but melting points are easier to realize, and
the difference in uncertainty (less than 2 mK for most
applications) is generally insignificant. In fact, the difference
between using traditional cells at their freezing points and
Hart’s mini cells at their melting points is not significant for
most labs in most applications.
Comparison blocks are also available for the 9260 for
high-precision comparison calibrations at high temperatures. Two
blocks are available with a variety of pre-drilled wells in
addition to blank or custom blocks. Well depth is 229 mm.
Features
- Good introduction to fixed-point calibration
- User friendly and inexpensive
|
|
|
Specifications |
|
Temperature Range |
50°C to 680°C |
|
Ambient Operating Range |
5°C to 45°C
|
|
Stability |
±0.03°C to 300°C
, ±0.05°C above 300°C |
|
Vertical Gradient |
Top and bottom
zones adjustable by offset |
|
Plateau Duration |
6 to 10 hours
typical |
|
Resolution |
0.01° |
|
Display Scale |
°C or °F,
switchable |
|
Immersion Depth |
229mm |
|
Stabilization Time |
15 minutes nominal |
|
Preheat Wells |
2 |
|
Fault Protection |
Sensor burnout and short protection,
over-temperature thermal cutout. |
|
Display Accuracy |
±0.2°C to 300°C, ±0.3°C
to 450°C, ±0.5°C to 680°C |
|
Comparison Block |
Two multi-hole blocks, blanks, and
custom blocks available |
|
Well-to-well Gradient
(in comparison block) |
±0,02°C |
|
Heating Time |
1.25 hours from 25°C
to 680°C |
|
Cooling Time |
10.5 hours from 680°C
to 100°C |
|
Communication Port |
RS-232 included |
|
Power Requirements |
115Vac (±10%),
60Hz, 11A; or, 230Vac (±10%) 50Hz, 6A, 1200W |
|
Exterior Dimensions |
489mm (H) x 222mm (W) x 260mm (D) |
|
Weight |
20.5kg |
|
|
|
|
|
|
Ordering Information |
|
Hart 9260 |
Mini Fixed-Point Furnace |
|
|
|
|
|
|
Accessories |
|
Hart 5914A |
Mini Quartz Indium Cell |
|
Hart 5915A |
Mini Quartz Tin Cell |
|
Hart 5916A |
Mini Quartz Zinc Cell |
|
Hart 5917A |
Mini Quartz Aluminum Cell |
|
Hart 5944 |
Metal Cased Mini Indium Cell |
|
Hart 5945 |
Metal Cased Mini Tin Cell |
|
Hart 5946 |
Metal Cased Mini Zinc Cell |
|
Hart 5947 |
Metal Cased Mini Aluminum Cell |
|
Hart 2940-9260 |
Container, Mini-Cell Support, for 9260 |
|
Hart 2942-9260 |
Container, SST Mini-Cell Support, for
9260 |
|
Hart 1904-X |
Accredited Cell Intercomparison |
|
Hart 3160-1 |
Comparison Insert, Blank |
|
Hart 3160-2 |
Comparison Insert, 7 holes at 6.35mm |
|
Hart 3160-3
|
Comparison Insert, 2 holes at 3.2mm, 2
holes at 4.76mm, 2 holes at 6.35mm,
2 holes at 9mm, 2 holes at 9.5mm |
|
|
Call for other comparison insert options |
|
|