DLRO10HD Dual Power 10A Low Resistance Ohmmeter - Specifications

Resistance/Current Ranges
The green resistance ranges on the keypad indicate low output power (<0.25W) outputs. Red ranges indicate higher output power 2.5W (1A) and 25W (10A).

Resolution and Accuracy
Test current accuracy ±10%
Voltmeter input impedance >200kΩ
 

Test Current
 
Resistance Range
 
Resolution
 
Basic Accuracy*
 
Full Scale
 Voltage
Maximum Power
Output

0.1mA

0 to 2500.0Ω

100mΩ

±0.2% ±200mΩ

25mV

2.5µW

0.1mA

0 to 250.00Ω

10mΩ

±0.2% ±20mΩ

25mV

2.5µW

1mA

0 to 25.000Ω

1mΩ

±0.2% ±2mΩ

25mV

25µW

10mA

0 to 2500.0

100µΩ

±0.2% ±200µΩ

25mV

250µW

100mA

0 to 250.00

10µΩ

±0.2% ±20µΩ

25mV

2.5mW

1A

0 to 25.000

1µΩ

±0.2% ±2µΩ

25mV

25mW

10A

0 to 2500.0

0.1µΩ

±0.2% ±0.2µΩ

25mV

0.25W

1A

0 to 2500.0

100µΩ

±0.2% ±20µΩ

2.5V

2.5W

10A

0 to 250.00

10µΩ

±0.2% ±20µΩ

2.5V

25W

 

Note*:

The accuracy stated assumes forward and reverse measurements.
  Inductive mode or unidirectional mode will induce an undefined error if an external EMF is present.
  Basic accuracy at reference conditions.
 

Specification

Display Type

Main:

5 digit

Secondary:

2 x 5 digit

Backlight

LED backlight

Auto Power Down

Powers down after 5 minutes

Range Selection Mode

Rotary switch

Noise Rejection

Less than 1% ±20 digits additional error with 100mV peak 50/60Hz on the potential leads. Warning will show if hum or noise exceeds this level

Maximum Lead Resistance

100mΩ total for 10 Amp operation irrespective of battery condition

 
 

Environmental

Temperature Coefficient

<0.01% per °C from 5°C to 40°C

Operating Temperature

-10°C to +50°C

Storage Temperature

-25°C to +60°C R.H. <90%

Humidity

<90% R.H. (reference conditions 20°C ±3°C

Maximum Altitude

2000 meters to full safety specifications.

 
 

Safety

Safety Rating

In accordance with IEC 61010-1 CAT III 300V when used with DH7 leads

IP Rating

IP65 case closed, IP54 with battery operation

EMC

In accordance with IEC 61326-1 (Heavy Industrial)

 
 

General Specification

Power Supply Input Voltage

90V to 264V 50/60Hz

Battery Type

6 volt 7AH, sealed lead acid battery.

Battery Life

1000 automatic three second tests.

Battery Charge Time

Eight hours

Dimensions

315mm (L) x 285mm (W) x 181mm (D)

Weight

6.7kg

 