|
Specifications |
|
Measurement Range |
0.1µΩ to 999.9mΩ
(subject to supply voltage and test leads used) |
|
Accuracy |
Voltage |
|
Current |
|
Resistance |
|
|
|
|
±0.5%
±0.1mV |
|
±0.5% ±0.1A |
|
Better than 1% from
300µΩ to 100mΩ |
|
Better than
±2% from 100µΩ to 300µΩ |
|
|
Current Lead Resistance
(Megger supplied leads) |
|
Resistance 8mΩ |
2 x 5 meter (L) 25mm2
(D) current leads |
|
Resistance 4mΩ |
2 x 5 meter (L) 50mm2
(D) current leads |
|
Resistance 5.4mΩ |
2 x 5 meter (L) 70mm2
(D) current leads |
|
Resistance 6mΩ |
2 x 5 meter (L) 95mm2
(D) current leads |
|
|
Maximum Continuous Test Time |
More than 60 seconds at 200 Amps @ 20°C
ambient |
|
Test Current Range |
10 Amp to 200 Amp DC |
|
Test Current Accuracy |
±2% ±2A |
|
Voltmeter Input Impedance |
> 200kΩ |
|
Hum Rejection |
5 Vrms 50/60Hz |
|
Test Modes |
Manual, Auto, Continuous |
|
Test Time |
10 seconds NORMAL/AUTO mode, refreshed
every 2-seconds in CONTINUOUS mode |
|
Display |
Large, high resolution backlit liquid
crystal display |
|
Warnings |
Current flowing: LED. Other
warnings: are shown on the LED display |
|
Data Transfer |
Real time or batch download via RS-232
using Download Manager |
|
Storage Capacity |
300 result sets and memo, battery
backed for 10 years |
|
Maximum Field Size |
160 characters maximum |