MIT200
Series Insulation Tester - Specifications |
|
|
Insulation Ranges |
|
Nominal Test Voltage |
250V,
500V, 1000V (DC) |
|
Measuring Ranges |
10kΩ to 1000MΩ on all ranges |
|
Terminal Voltage open circuit (dc) |
-0% + 25% of rated voltage |
|
Short Circuit Current |
205mA +10mA -5mA |
Test Current on Load
|
>1mA at maximum pass values of insulation as specified in
BS 7671, HD 384, IEC 364, and VDE 0413 part 1 |
Accuracy (at 20°C)
|
±3% of reading ±2 digits up to 10MΩ
±5% of reading ±2 digits up to 100MΩ
±30% of reading up to 1000MΩ |
for Models
MIT200, MIT210, MIT220, MIT230 |
Terminal Characteristics
 |
|
|
|
|
|
|
Continuity Ranges |
|
Measuring Range |
0.01Ω to 100.0Ω (0Ω to 50Ω on analog scale) |
|
Open Circuit Voltage |
5V ±1V |
|
Short Circuit Current |
205mA ± 5mA (0Ω to 10Ω) > 20mA (0Ω to 100Ω) Ω |
Accuracy (at 20°C)
|
±0.01Ω to 9.99Ω ±5% ±2 digits
±10.0Ω to 99.9Ω ±5% ±2 digits |
for Models
MIT200, MIT210, MIT220, MIT230 |
|
Zero Offset Adjust |
0Ω to 9.99Ω |
|
Continuity Buzzer |
Operates at < 2Ω |
|
|
|
|
|
|
Specifications |
|
Default Voltmeter |
> 25V AC or DC is applied display will operate as a voltmeter |
|
Test Inhibit |
If more than 50 volts is detected, testing will be inhibited |
|
Voltmeter Range |
25V to 600V at 50/60Hz and DC |
|
Voltmeter Accuracy |
25V to 450V AC/DC (±1% ± 1 digit)
450V to 600V AC/DC (±2% ± 1 digit) |
|
Auto Power Down |
Operates after 10 minutes if left in standby mode |
Automatic Discharge
|
After an insulation test the item under test will be discharged
automatically. Any voltage present will be indicated on the display
so that the discharge can be monitored. |
|
|
|
|
|
|
Environmental Specifications |
|
Operating Temperature |
-10°C to +55°C |
|
Storage Temperature |
-25°C to +65°C |
|
Operating Humidity |
93% R.H. at +40°C maximum |
|
EMC |
In accordance with IEC 61326-1 |
|
Environmental Protection |
IP40 |
|
|
|
|
|
|
Safety |
|
Meets the requirements |
of
EN 61010-1 CAT III 600V, phase to earth |
|
Terminal Fuse |
500mA (fast blow) 600V, 32mm x 6mm, ceramic HBC 50kA minimum. |
Automatic Discharge
|
After an insulation test the item under test will be discharged
automatically. Any voltage present will be indicated on the display
so that the discharge can be monitored. |
|
|
|
|
|
|
General Specifications |
|
Power Supply (battery) |
Six 1.5V Alkaline type 'AA' (LR6) batteries
(Rechargeable NiMH cells may be used) |
|
Battery Life |
3000 consecutive tests (5 seconds per test) on any test using 2Ah
batteries |
|
Dimensions |
195mm (L) x 98mm (W) x 40mm (D) |
|
Weight |
550g approx. |
|